MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.70839

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026

Comparison Between the Acurate Neo and the Acurate Neo 2 Transcatheter Heart Valve System: Impact of Midframe Under-Expansion.

Crusius Lisa, Ast Thaddäus, Haussig Stephan, Steul Jean-Honoré, Woitek Felix J, Sick Peter, Madej Tomas, Kappert Utz, Linke Axel, Mangner Norman

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Transcatheter aortic valve implantation (TAVI) with the Acurate Neo has been linked to a significant incidence of paravalvular leakage (PVL), addressed by the Acurate Neo 2 incorporating an active sealing skirt. Recent findings suggest that midframe under-expansion is associated with worse outcomes.

AIMS: We investigated the hemodynamic and clinical outcomes after TAVI with Acurate Neo 2 compared to Acurate Neo and assessed the impact of midframe under-expansion.

METHODS

Patients who received Acurate Neo (n = 124) or Acurate Neo 2 (n = 133) between February 2018 and December 2021 (switch to Acurate Neo 2 in November 2020) were included. The primary outcomes were VARC-3 defined device success and early safety.

RESULTS

Device success was higher with Acurate Neo 2 compared with Acurate Neo (94.7% vs. 86.3%, p = 0.020), as was early safety (87.2% vs. 74.2%, p = 0.008). The rate of mild/moderate PVL was significantly lower in Acurate Neo 2. Kaplan-Meier estimated 1-year survival (Acurate Neo 2: 87.8% vs. Acurate Neo: 89.0, p = 0.930) and freedom from the combined clinical endpoint death, stroke, myocardial infarction, and heart failure rehospitalization at 1 year (Acurate Neo 2: 81.2% vs. Acurate Neo: 79.3%, p = 0.528) were comparable between groups. A long-term analysis of these endpoints revealed no difference between groups. Midframe under-expansion occurred in 13.3% and was associated with lower freedom from the combined clinical endpoint at 1 year, irrespective of prosthesis generation.

CONCLUSION

Acurate Neo 2 demonstrated significant improvements in device success and early safety compared to Acurate Neo. Midframe under-expansion occurred with both generations and was associated with worse clinical outcomes within the first year.

Для цитирования: Crusius Lisa, Ast Thaddäus, Haussig Stephan, Steul Jean-Honoré, Woitek Felix J, Sick Peter, Madej Tomas, Kappert Utz, Linke Axel, Mangner Norman. Comparison Between the Acurate Neo and the Acurate Neo 2 Transcatheter Heart Valve System: Impact of Midframe Under-Expansion.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2026. DOI: 10.1002/ccd.70839.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.