BACKGROUND

Rapid evaluation for pneumothorax is a common clinical priority. Although lung ultrasound (LUS) often is used to assess for pneumothorax, its diagnostic accuracy varies based on patient and provider factors. To enhance the performance of LUS for pulmonary pathologic features, artificial intelligence (AI)-assisted imaging has been adopted; however, the diagnostic accuracy of AI-assisted LUS (AI-LUS) deployed in real time to diagnose pneumothorax remains unknown.

RESEARCH QUESTION: In patients with suspected pneumothorax, what is the real-time diagnostic accuracy of AI-LUS to recognize the absence of lung sliding?

STUDY DESIGN AND

METHODS

We performed a prospective AI-assisted diagnostic accuracy study of AI-LUS to recognize the absence of lung sliding in a convenience sample of patients with suspected pneumothorax. After calibrating the model parameters and imaging settings for bedside deployment, we prospectively evaluated its diagnostic accuracy for lung sliding compared with a reference standard of expert consensus.

RESULTS

Two hundred forty-one lung sliding evaluations were derived from 62 patients. AI-LUS showed a sensitivity of 0.921 (95% CI, 0.792-0.973), specificity of 0.802 (95% CI, 0.735-0.856), area under the receiver operating characteristic curve of 0.885 (95% CI, 0.828-0.956), and accuracy of 0.824 (95% CI, 0.766-0.870) for the diagnosis of absent lung sliding.

INTERPRETATION: In this study, real-time AI-LUS showed high sensitivity and moderate specificity to identify the absence of lung sliding. Further research to improve model performance and optimize the integration of AI-LUS into existing diagnostic pathways is warranted.

Для цитирования: Clausdorff Fiedler Hans, Prager Ross, Smith Delaney, Wu Derek, Dave Chintan, Tschirhart Jared, Wu Ben, Van Berlo Blake, Malthaner Richard, Arntfield Robert. Automated Real-Time Detection of Lung Sliding Using Artificial Intelligence: A Prospective Diagnostic Accuracy Study.. Chest. 2024. DOI: 10.1016/j.chest.2024.02.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.