Аннотация

PURPOSE: We propose a new motion tracking method that encodes an object's position information using tagging magnetic resonance (MR) images as digital codes, and we assess the method's feasibility in stationary and moving phantoms. The encoding and decoding of tag patterns employ principles of spread spectrum communication.

METHODS

We used a segmented fast low angle shot cine sequence (FLASH) with spatial modulation of magnetization (SPAMM) preparation pulses to encode position information as 7-bit code words and used this spread code to decode the position information. To make 7-bit code words using tag images, we adjusted the flip angle and phase of the 4 composite radiofrequency (RF) pulses and the gradient magnetic field strength in the SPAMM pulse to generate 7 types of tag patterns. The proposed method required 7 acquisitions with 7 types of tag patterns. We compared results with images obtained by conventional tagging in stationary and moving phantom experiments.

RESULTS

In a stationary phantom, the proposed method showed the same ability to identify pixel position as conventional tagging method using improved SNR images with the average of 7 acquisitions. In a moving phantom, pixel position was successfully decoded by the proposed method on a pixel-by-pixel basis. In this method, the motion of the phantom was detected by the simple calculation of the correlation coefficient of the code words.

CONCLUSION

We introduced a spread spectrum communication technique to tagging MR imaging that regards tag patterns as digital codes, and we demonstrated the method's potential to detect pixel position in sub-pixel resolution.

Для цитирования: Komori Yoshiaki, Amano Akira, Maehara Keiko, Jin Li, Narazaki Michiko, Matsuda Tetsuya. A motion tracking method that applies a spread spectrum communication technique to tagging MR imaging.. Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine. 2014. DOI: 10.2463/mrms.2013-0116.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.