DOI 10.1016/j.jcrc.2020.10.005
Automatic oxygen administration and weaning in patients following mechanical ventilation.
Аннотация
PURPOSE: To evaluate efficacy of FreeO2 device in oxygen weaning of patients after being liberated from mechanical ventilation (MV).
METHODS
Prospective crossover cohort study in patients admitted to ICU and after MV weaning. FreeO2 curves were recorded during constant flow and FreeO2 modes. Oxygenation parameters and O2 consumption were assessed.
RESULTS
Fifty one records were obtained in 51 patients (median age, 62 years, 54.9% had COPD, admission for acute respiratory failure in 96%). NIV was used initially in 68.6%. For a median records duration of 2.04 h, the time spent within target SpO2 range was significantly higher with FreeO2 mode compared to constant O2 flow mode [86.92% (77.11-92.39) vs 43.17% (5.08-75.37); p < 0.001]. Time with hyperoxia was lower with FreeO2 mode: 8.68% (2.96-15.59) vs 38.28% (2.02-86.34). Times with hypoxaemia, and with severe desaturation, were similar. At the end of FreeO2 mode, O2 flow was lower than 1 l/min in 28 patients (54.9%), with a median of 0.99 l/min.
CONCLUSIONS
For the purpose of oxygen weaning in patients recovering from MV, automatic O2 titration with FreeO2 was associated with a substantial reduction in O2 delivery and better oxygenation parameters in comparison with constant O2 flow.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.