MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00345-019-02808-5

Медицинская библиотека · World journal of urology · 2019

Thermal effects of Ho:YAG laser lithotripsy during retrograde intrarenal surgery and percutaneous nephrolithotomy in an ex vivo porcine kidney model.

Hein Simon, Petzold Ralf, Suarez-Ibarrola Rodrigo, Müller Philippe-Fabian, Schoenthaler Martin, Miernik Arkadiusz

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the thermal effect of high-power holmium:yttrium-aluminum-garnet (Ho:YAG) laser lithotripsy in flexible/semirigid ureteroscopy (fURS/sURS) and percutaneous nephrolithotomy (PNL) in a standardized ex vivo porcine kidney model with real-time temperature assessment.

METHODS

The experimental setup consisted of three models designed to evaluate the thermal effects of Ho:YAG laser lithotripsy in fURS, sURS and PNL, respectively. In all setups, a postmortem porcine kidney was placed in a 37 °C water bath. Three thermocouples were inserted into the renal parenchyma while a flexible thermocouple was placed 3-4 mm proximal to the laser fiber to measure temperature variations in the collecting system. The thermal impact was evaluated in relation to laser power between 5 and 100 W and various irrigation rates (37 °C, 0-100 ml/min).

RESULTS

In all three experimental setups, sufficient irrigation was required to prevent potentially damaging temperatures into the renal pelvis and parenchyma. Even 5 W in fURS can lead to a potentially harming temperature rise if insufficient irrigation is applied. Particularly, high-power settings ≥ 30 W carry an elevated risk for critical temperature rises. The results allow the definition of a specific irrigation threshold for any power setting to prevent critical temperatures in the present study design.

CONCLUSIONS

Ho:YAG laser lithotripsy bears the risk of thermal damages to the urinary tract even at low-power settings if inadequate irrigation is applied. Sufficient irrigation is mandatory to perform safe Ho:YAG laser lithotripsy. Based on the results, we developed a formula calculating the approximate ΔT for irrigation rates ≥ 30 ml/min: ΔT = 15 K × (power [W]/irrigation [ml/min]).

Для цитирования: Hein Simon, Petzold Ralf, Suarez-Ibarrola Rodrigo, Müller Philippe-Fabian, Schoenthaler Martin, Miernik Arkadiusz. Thermal effects of Ho:YAG laser lithotripsy during retrograde intrarenal surgery and percutaneous nephrolithotomy in an ex vivo porcine kidney model.. World journal of urology. 2019. DOI: 10.1007/s00345-019-02808-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.