The Rise and Fall of High Temperatures During Ureteroscopic Holmium Laser Lithotripsy.

Winship Brenton, Wollin Daniel, Carlos Evan, Peters Chloe, Li Jingqiu, Terry Russell, Boydston Kohldon, Preminger Glenn M, Lipkin Michael E

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Introduction

Temperatures over 43°C-the threshold for cellular injury-may be achieved during ureteroscopic holmium laser lithotripsy. The time to reach and subsequently clear high temperatures at variable laser power settings and irrigation pressures has not been studied.

Methods

A flexible or semirigid ureteroscope was placed within an 11/13 F ureteral access sheath inserted into a 250-mL saline bag simulating a normal-caliber ureter, renal pelvis reservoir, and antegrade irrigation flow. A thermocouple was placed adjacent to a 365 μm fiber fired for 45 seconds at 0.6 J/6 Hz, 0.8 J/8 Hz, 1 J/10 Hz, 1 J/20 Hz, and 0.2 J/80 Hz. Irrigation pressures of 200, 100, and 0 mm Hg were tested. Mean temperature changes were recorded with 6°C increase as a threshold for injury (as body temperature is 6°C below 43°C).

Results

Semirigid scope: At 200 mm Hg no temperature changes >6°C were observed. At 100 mm Hg, changes >6°C occurred with 1 J/20 Hz within 1 second of activation and returned to ≤6°C within 1 second of cessation. At 0 mm Hg, changes >6°C occurred with all settings; within 1 second at power ≥10 W. Temperatures returned to ≤6°C within 5-10 seconds. Flexible scope: At 200 mm Hg, changes >6°C occurred at 1 J/10 Hz (15 seconds), 0.2 J/80 Hz (3 seconds), and 1 J/20 Hz (2 seconds). Temperatures returned within 6°C of baseline within 2 seconds. At 100 mm Hg, changes >6°C occurred in all but 0.6 J/6 Hz. Temperatures returned to ≤6°C in 5-10 seconds. At 0 mm Hg, all settings produced changes >6°C within 3 seconds, except 0.6 J/6 Hz (35 seconds). Temperatures returned to ≤6°C in under 10 seconds.

Conclusions

High temperatures were achieved in our in vitro model in as little as 1 second at common irrigation pressures and laser settings, particularly with a flexible ureteroscope and power ≥10 W. However, with laser cessation, temperatures quickly returned to a safe level at each irrigation pressure.

Для цитирования: Winship Brenton, Wollin Daniel, Carlos Evan, Peters Chloe, Li Jingqiu, Terry Russell, Boydston Kohldon, Preminger Glenn M, Lipkin Michael E. The Rise and Fall of High Temperatures During Ureteroscopic Holmium Laser Lithotripsy.. Journal of endourology. 2019. DOI: 10.1089/end.2019.0084.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.