DOI 10.1038/s41433-023-02528-6
Repeatability of Scheimpflug based corneal tomography parameters in advanced keratoconus with thin corneas.
Аннотация
OBJECTIVE
To determine the repeatability limits of corneal tomography parameters in patients with advanced and moderately thin keratoconic corneas to assist in planning thickness-based procedural interventions.
METHODS
Prospective, single-centre, repeatability study. Three tomography scans using the Pentacam AXL were obtained from patients with keratoconus with thinnest corneal thickness (TCT) ≦400 µm (sub-400 group) and compared to those with TCT = 450-500 µm (450-plus group). Eyes with previous crosslinking, intraocular surgery, or acute corneal hydrops were excluded. Eyes were age and gender-matched. The within-subject standard deviations for flat keratometry (K1), steep keratometry (K2), maximal keratometry (Kmax), astigmatism and TCT were used to calculate respective repeatability limits (r). Intra-class correlation coefficients (ICC) were also analysed.
RESULTS
The sub-400 group comprised 114 eyes from 114 participants, and the 450-plus group comprised 114 eyes from 114 participants. In the sub-400 group, TCT was amongst the least repeatable parameters (33.92 µm; ICC 0.96), compared with the 450-plus group (14.32 µm; ICC 0.99, p < 0.01). In the sub-400 group, K1 and K2 of the anterior surface were the most repeatable parameters (r 3.79 and 3.22 respectively; ICC 0.97 and 0.98 respectively) compared with the 450-plus group (r 1.17 and 0.92 respectively; and ICC 0.98 and 0.99 respectively, p < 0.01).
CONCLUSIONS
The repeatability of corneal tomography measurements is significantly reduced in sub-400 keratoconic corneas when compared to 450-plus corneas. Repeatability limits should be carefully considered when surgical interventions are planned for such patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.