DOI 10.1007/s10792-023-02693-3

Медицинская библиотека · International ophthalmology · 2023

Measuring the agreement of keratometry readings of four devices in eyes with keratoconus.

Abdi Saeid, Hosseini Abardeh Mohsen, Naseri Mohammad Hashem, Kamary Rad Zahra, Naghdi Taghi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the keratometry measurement agreement using Sirius corneal topography and Scheimpflug camera, Tomey corneal topography, Topcon autokeratorefractometer, and Tomey OA-2000 optical biometry in eyes with different severity of keratoconus.

METHODS

In this retrospective study 115 eyes in different stages of keratoconus were divided into 2 groups of mild (stage 1), and moderate to severe keratoconus (stages 2, 3), according to the Amsler-Krumeich classification. Keratometry measurements were obtained using Sirius corneal topography and Scheimpflug camera (phoenix V3.7.01.08), Tomey corneal topography (Tms SW22C-200S-200), Topcon autokeratorefractometer (KR8900), and Tomey optical biometry (OA-2000 Opt-Meas V.4E).

RESULTS

In group 1 All devices demonstrated fair agreement in average keratometry values (95% LoA range > 1 D). However, it was poorer for group 2 (95% LOA range > 3 D). In group 1 Bonferroni test revealed statistically significant difference in average K readings among (Topcon autokeratorefractometer 8900 and Tomey OA-2000 biometry, p < 0.01), and between (Tomey topography and Topcon autokeratorefractometer, p < 0.05). Also in group 2 significant difference was observed in the average keratometry of the most instruments (p < 0.01) except for the (Tomey topography and Topcon autokeratorefractometer) and (Tomey topography and Tomey OA-2000 biometry.

CONCLUSIONS

According to our investigation in mild, moderate and severe keratoconus the agreement in K reading between Topcon autokeratorefractometer, OA-2000 optical biometry, Sirius topography and Tomey topography was poor. The agreement declines especially in the steep meridian and it was not acceptable clinically. These devices should not be applied interchangeably.

Для цитирования: Abdi Saeid, Hosseini Abardeh Mohsen, Naseri Mohammad Hashem, Kamary Rad Zahra, Naghdi Taghi. Measuring the agreement of keratometry readings of four devices in eyes with keratoconus.. International ophthalmology. 2023. DOI: 10.1007/s10792-023-02693-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.