Motion-Tracking Brillouin Microscopy Evaluation of Normal, Keratoconic, and Post-Laser Vision Correction Corneas.

Zhang Hongyuan, Asroui Lara, Tarib Imane, Dupps William J, Scarcelli Giuliano, Randleman J Bradley

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

PURPOSE: To characterize focal biomechanical differences between normal, keratoconic, and post-laser vision correction (LVC) corneas using motion-tracking Brillouin microscopy.

DESIGN

Prospective cross-sectional study.

METHODS

Thirty eyes from 30 patients (10 normal controls [Controls], 10 post-LVC, and 10 stage I or II keratoconus [KC]) had Scheimpflug and motion-tracking Brillouin microscopy imaging using a custom-built device. Mean, maximum (max) and minimum (min) Brillouin shift, spatial standard deviation, and max-min values were compared. Min values were correlated with local Brillouin values at multiple Scheimpflug imaging locations.

RESULTS

Mean (P < .0003), min (P < .00001), spatial standard deviation (P < .01), and max-min (P < .001) were significantly different between the groups. In post hoc pairwise comparisons, the best differentiators for group comparisons were mean (P = .0004) and min (P = .000002) for Controls vs KC, min (P = .0022) and max-min (P = .002) for Controls vs LVC, and mean (P = .0037) and min (P = .0043) for LVC vs KC. Min (area under the receiver operating characteristic = 1.0) and mean (area under the receiver operating characteristic =  0.96) performed well in differentiating Control and KC eyes. Min values correlated best with Brillouin shift values at the thinnest corneal point (r2 = 0.871, P = .001) and maximum keratometry value identified in the tangential curvature map (r2 = 0.840, P = .002).

CONCLUSIONS

Motion-tracking Brillouin microscopy effectively characterized focal corneal biomechanical alterations in LVC and KC and clearly differentiated these groups from Controls. Primary motion-tracking Brillouin metrics performed well in differentiating groups as compared with basic Scheimpflug metrics, in contrast to previous Brillouin studies, and identified focal changes after LVC where prior Brillouin studies did not.

Для цитирования: Zhang Hongyuan, Asroui Lara, Tarib Imane, Dupps William J, Scarcelli Giuliano, Randleman J Bradley. Motion-Tracking Brillouin Microscopy Evaluation of Normal, Keratoconic, and Post-Laser Vision Correction Corneas.. American journal of ophthalmology. 2023. DOI: 10.1016/j.ajo.2023.03.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.