DOI 10.4103/ojo.ojo_147_21

Медицинская библиотека · Oman journal of ophthalmology · 2022

Comparison of IOP obtained in different kind of eyes with contact and no contact tonometers.

Sbordone Sandro, Ragucci Adele, Iaccarino Gennarfrancesco, Scognamiglio Gabriele, Leone Angelo, Carnevale Ugo Antonello Gironi, Lanza Michele

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The aim of this study is to analyze and compare the intraocular pressure (IOP) values measured in three different kinds of participants such as healthy subjects (HS), keratoconus patients (KP), and those who underwent myopic photorefractive keratectomy (MPRK). The devices used in this study are the Goldmann Applanation Tonometry (GAT), the dynamic contour tonometry (DCT), the ocular response analyzer (ORA), and the Corvis ST (CST).

SUBJECTS AND

METHODS

This research included 92 eyes of 92 h, 63 eyes of 63 KP, and 58 eyes of 58 MPRKM. Each participant underwent a complete ophthalmic evaluation and IOP measurement with GAT, DCT, ORA, and CST. A statistical analysis was conducted to detect possible differences and correlations.

RESULTS

First, according to the observed data, HS eyes displayed mean IOP values measured with GAT, DCT, ORA, and CST, respectively, 15.82 ± 2.74 mmHg, 17.63 ± 2.28 mmHg, 16.24 ± 3.14 mmHg, and 17.31 ± 3.21 mmHg. Then, KP eyes showed mean IOP values measured with GAT, DCT, ORA, and CST of, respectively, 14.89 ± 1.64 mmHg, 16.97 ± 2.08 mmHg, 13.09 ± 3.12 mmHg, and 13.78 ± 2.11 mmHg. Finally, MPRK eyes showed mean IOP values measured with GAT, DCT, ORA, and CST of, respectively, 13.92 ± 1.34 mmHg, 15.39 ± 2.86 mmHg, 16.63 ± 2.51 mmHg, and 15.06 ± 1.56 mmHg.

CONCLUSION

According to the observed data, ORA and GAT might be used interchangeably in HS, whereas GAT, ORA, and CST in KP eyes. Moreover, it has been noticed that in those eyes that previously undergone a myopic PRK, GAT provided lower values of IOP in comparison with other devices.

Для цитирования: Sbordone Sandro, Ragucci Adele, Iaccarino Gennarfrancesco, Scognamiglio Gabriele, Leone Angelo, Carnevale Ugo Antonello Gironi, Lanza Michele. Comparison of IOP obtained in different kind of eyes with contact and no contact tonometers.. Oman journal of ophthalmology. 2022. DOI: 10.4103/ojo.ojo_147_21.