DOI 10.1097/j.jcrs.0000000000001136

Медицинская библиотека · Journal of cataract and refractive surgery · 2023

Repeatability of corneal deformation response parameters by dynamic ultra-high-speed Scheimpflug imaging before and after corneal crosslinking.

Herber Robert, Vinciguerra Riccardo, Tredici Costanza, Legrottaglie Emanuela F, Pillunat Lutz E, Raiskup Frederik, Vinciguerra Paolo

Оценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the repeatability of deformation corneal response (DCR) parameters before and after corneal crosslinking (CXL) compared with their untreated fellow eyes (uFEs).

SETTING

University Hospital Carl Gustav Carus, Dresden, Germany; IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.

DESIGN

Multicenter, interventional reliability analysis.

METHODS

53 eyes of 53 patients with keratoconus who received CXL treatment after the disease progression (CXL group) were included. Patients were measured 3 times using a dynamic Scheimpflug analyzer to determine repeatability before and 1 month after CXL treatment. The uFEs were measured in the same way (uFE group). Reliability of DCR parameters was assessed by a coefficient of repeatability, coefficient of variation, and intraclass correlation coefficient (ICC).

RESULTS

The repeatability of DCR parameters did not change after CXL compared with the preoperative values for all investigated DCR parameters ( P > .05). In the uFE group, no statistically significant shift was observed regarding the repeatability ( P > .05). An ICC greater than 0.75 was achieved in both groups for almost all parameters. Concerning the biomechanical stiffening induced by CXL, integrated inverse radius and stress-strain index were found to be statistically significantly decreased and increased ( P < .001), respectively, both indicating stiffening. No changes were observed for the uFE group.

CONCLUSIONS

The study demonstrated highly repeatable measurements of the dynamic Scheimpflug analyzer before and after CXL. The improvement of certain DCR parameters after CXL confirmed the capability of the device to detect the stiffening effect.

Для цитирования: Herber Robert, Vinciguerra Riccardo, Tredici Costanza, Legrottaglie Emanuela F, Pillunat Lutz E, Raiskup Frederik, Vinciguerra Paolo. Repeatability of corneal deformation response parameters by dynamic ultra-high-speed Scheimpflug imaging before and after corneal crosslinking.. Journal of cataract and refractive surgery. 2023. DOI: 10.1097/j.jcrs.0000000000001136.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.