Single-centre experience on endocardial and epicardial pacemaker system function in neonates and infants.

Silvetti Massimo Stefano, Drago Fabrizio, De Santis Antonella, Grutter Giorgia, Ravà Lucilla, Monti Lidia, Fruhwirth Rodolfo

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Endocardial (ENDO) or epicardial (EPI) pacing systems are implanted in infants but it remains unclear which system should be preferred.

METHODS AND

RESULTS

We evaluated the results of children <or=1 year who underwent pacemaker (PM) implantation at our centre with a retrospective analysis. Between 1992 and 2004, 56 patients, 37 of whom had other congenital heart defects (CHDs), received a PM at 4.4 +/- 3.8 months of age for atrioventricular block (n = 52) and sinus node dysfunction. Rate-responsive ventricular demand pacing (VVIR) PMs were implanted in 25 patients (19 ENDO), dual-chamber demand pacing (DDD) in 29, and rate-responsive atrial demand pacing (AAIR) in 2 (all EPI). Follow-up (FU) was 4.5 +/- 3.5 (range 0.3-13) years: 15 pacing system failures occurred among the 56 patients (26%) after 4.5 +/- 3.2 years, with a significantly reduced success rate for EPI (21-fold increase of the risk of failure) and complex CHD. Also in patients without surgery for CHD, EPI showed a worse outcome. Among the 91 leads implanted, failures occurred more significantly in EPI (18% of atrial, 24% of ventricular leads) than in ENDO (5% of ventricular leads). No venous occlusion was found at FU.

CONCLUSIONS

Single-lead, VVIR ENDO pacing had higher efficiency and safety than EPI, and it might be the best choice for PM implantation in infants. However, because of small patient numbers and lack of longer FU, these findings should be treated with caution.

Для цитирования: Silvetti Massimo Stefano, Drago Fabrizio, De Santis Antonella, Grutter Giorgia, Ravà Lucilla, Monti Lidia, Fruhwirth Rodolfo. Single-centre experience on endocardial and epicardial pacemaker system function in neonates and infants.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2007. DOI: 10.1093/europace/eum043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.