DOI 10.1111/j.1540-8159.2005.40054.x
Rate-dependent AV delay optimization in cardiac resynchronization therapy.
Аннотация
BACKGROUND
During cardiac resynchronization therapy (CRT), cardiac performance is dependent on an optimized atrioventricular delay (AVD). However, the optimal AVD at different heart rates has not been defined yet during CRT.
METHOD: The effects of an increase in heart rate by pacing or physical exercise on optimal AVD were studied in 36 patients with biventricular pacemakers/defibrillators. The velocity time integral (VTI) in the left ventricular outflow tract (LVOT) was measured with pulsed Doppler either at three different paced heart rates in the supine position or in seated position before and after physical exercise.
RESULTS
The baseline AVD was optimized to 99 +/- 19 ms in the supine and 84 +/- 22 ms in the seated position. When the heart rate was increased by DDD pacing, there was a positive linear relationship between an increase in heart rate, in AVD and in VTI (LVOT-VTI + 0.047 cm/s per 10 beats per minute (bpm) heart rate increase per 20 ms increase in AVD, P = 0.007). A similar but more pronounced relationship was found after physical exercise in the seated position (LVOT-VTI + 0.146 cm/s per 10 bpm heart rate increase per 20 ms increase of AVD, P = 0.013). This effect was observed in patients with and without AV block and mitral regurgitation.
CONCLUSIONS
In conclusion, the systolic performance of the dilated ventricle, which depends on an elevated preload, is critically affected by the appropriate timing of the AVD during exercise. In contrast to normal pacemaker patients, in CRT the relatively short baseline AVD should be prolonged at increased heart rates. Further studies with other means of measuring exercise cardiac performance are needed to confirm these unexpected findings.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.