MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/sleep/zsac103

Медицинская библиотека · Sleep · 2022

Hypocretin-1 measurements in cerebrospinal fluid using radioimmunoassay: within and between assay reliability and limit of quantification.

van der Hoeven Adrienne Elisabeth, van Waaij Kevin, Bijlenga Denise, Roelandse Frederik Willem Cornelis, Overeem Sebastiaan, Bakker Jaap Adriaan, Fronczek Rolf, Lammers Gert Jan

СтатьяПеревод пока не подготовлен

Аннотация

STUDY

OBJECTIVES

The most sensitive and specific investigative method for the diagnosis of narcolepsy type 1 (NT1) is the determination of hypocretin-1 (orexin-A) deficiency (≤110 pg/mL) in cerebrospinal fluid using a radioimmunoassay (RIA). We aimed to assess the reliability of the Phoenix Pharmaceuticals hypocretin-1 RIA, by determining the lower limit of quantification (LLOQ), the variability around the cutoff of 110 pg/mL, and the inter- and intra-assay variability.

METHODS

Raw data of 80 consecutive hypocretin-1 RIAs were used to estimate the intra- and inter-assay coefficient of variation (CV). The LLOQ was established and defined as the lowest converted concentration with a CV <25%; the conversion is performed using a harmonization sample which is internationally used to minimize variation between RIAs.

RESULTS

The mean intra-assay CV was 4.7%, while the unconverted inter-assay CV was 28.3% (18.5% excluding 2 outliers) and 7.5% when converted to international values. The LLOQ was determined as 27.9 pg/mL. The intra-assay CV of RIAs with lower specific radioactive activity showed a median of 5.6% (n = 41, range 1.6%-17.0%), which was significantly higher than in RIAs with higher specific activity (n = 36; median 3.2%, range 0.4%-11.6%, p = .013). The CV around the 110 pg/mL cutoff was <7%.

CONCLUSIONS

Hypocretin-1 RIAs should always be harmonized using standard reference material. The specific activity of an RIA has a significant impact on its reliability, because of the decay of 125I radioactivity. Values around the hypocretin-1 cut-off can reliably be measured. Hypocretin-1 concentrations below 28 pg/mL should be reported as "undetectable" when measured with the Phoenix Pharmaceuticals RIA.

CLINICAL TRIAL INFORMATION: This study is not registered in a clinical trial register, as it has a retrospective database design.

Для цитирования: van der Hoeven Adrienne Elisabeth, van Waaij Kevin, Bijlenga Denise, Roelandse Frederik Willem Cornelis, Overeem Sebastiaan, Bakker Jaap Adriaan, Fronczek Rolf, Lammers Gert Jan. Hypocretin-1 measurements in cerebrospinal fluid using radioimmunoassay: within and between assay reliability and limit of quantification.. Sleep. 2022. DOI: 10.1093/sleep/zsac103.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.