DOI 10.1111/j.1600-0536.2011.02035.x

Медицинская библиотека · Contact dermatitis · 2012

Stability of selected volatile contact allergens in different patch test chambers under different storage conditions.

Mose Kristian F, Andersen Klaus E, Christensen Lars Porskjaer

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Patch test preparations of volatile substances may evaporate during storage, thereby giving rise to reduced patch test concentrations.

OBJECTIVES

To investigate the stability of selected acrylates/methacrylates and fragrance allergens in three different test chambers under different storage conditions.

METHODS

Petrolatum samples of methyl methacrylate (MMA), 2-hydroxyethyl methacrylate (2-HEMA), 2-hydroxypropyl acrylate (2-HPA), cinnamal and eugenol in patch test concentrations were stored in three different test chambers (IQ chamber™, IQ Ultimate™, and Van der Bend® transport container) at room temperature and in a refrigerator. The samples were analysed in triplicate with high-performance liquid chromatography.

RESULTS

The decrease in concentration was substantial for all five allergens under both storage conditions in IQ chamber™ and IQ Ultimate™, with the exception of 2-HEMA during storage in the refrigerator. For these two chamber systems, the contact allergen concentration dropped below the stability limit in the following order: MMA, cinnamal, 2-HPA, eugenol, and 2-HEMA. In the Van der Bend® transport container, the contact allergens exhibited acceptable stability under both storage conditions, whereas MMA and 2-HPA required cool storage for maintenance of the limit.

CONCLUSION

The Van der Bend® transport container was the best device for storage of samples of volatile contact allergens.

Для цитирования: Mose Kristian F, Andersen Klaus E, Christensen Lars Porskjaer. Stability of selected volatile contact allergens in different patch test chambers under different storage conditions.. Contact dermatitis. 2012. DOI: 10.1111/j.1600-0536.2011.02035.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.