Qualitative assessment of tear dynamics with fluorescein profilometry.
Аннотация
PURPOSE: To describe a new methodology for tear-film dynamics assessment by observing fluorescein decay rate over time and to understand the relationship between the newly defined tear fluorescein washout rate (TFWR) and other measures of the tear film behaviour.
METHODS
Forty subjects (24F/16M) aged (mean±standard deviation) 31.8±14.2years volunteered for the study. It consisted of the review of medical history, McMonnies questionnaire (McMQ), slit lamp examination, and TFWR using a newly-developed fluorescein profilometry. The repeatability of TFWR measurements was assessed. TFWR estimates were contrasted against patient age, McMQ score, daytime, fluorescein tear film break-up time (FTBUT), tear meniscus height (TMH) and blink frequency.
RESULTS
Mean repeatability of the method was 28.13±9.59%. The group mean TFWR was 39±23% at 30-s mark after the beginning of measurements, ranging from 1.4% to 83%. This indicates that TFWR is highly subject-dependent. Statistically significant correlations were found between the percentage TFWR and McMQ score (r2=0.214, p=0.001) as well as FTBUT (r2=0.136, p=0.009). No statistically significant correlations were found between TFWR and age, daytime, TMH, and blink frequency.
CONCLUSIONS
Fluorescein profilometry allows clinicians to follow dynamic changes in the tear film on the entire ocular surface and may be used for qualitative assessment of the tear film dynamics.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.