MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euae213

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024

Extraction of lumenless pacing leads from the His bundle and left bundle branch area: outcomes of the high-volume centre.

Gardas Rafal, Loboda Danuta, Biernat Jolanta, Soral Tomasz, Kulesza Piotr, Gladysz-Wanha Sylwia, Joniec Michal, Sajdok Mateusz, Zub Kamil, Golba Krzysztof S

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: The application of conduction system pacing (CSP) in clinical practice is growing, and the need for lead extraction will also increase. The data on outcomes and safety of CSP lead extraction are limited. The aim of this study was to assess procedural outcomes and safety of CSP lead removal.

METHODS AND

RESULTS

Forty-seven patients from the EXTRACT Registry with the indication for CSP lead removal were enrolled in the study conducted at the Department of Electrocardiology in Katowice, Poland. Extraction technique, outcomes, safety, and complication were evaluated. Forty-three (91.5%) leads were successfully removed, and 41 (87.2%) were removed with traction only. The dwelling time of 28 extracted leads was longer than 1 year, and the oldest extracted lead was implanted for 89 months. Seven (14.9%) leads were removed from the left bundle branch (LBB) area and 36 from the His bundle (HB). Transient complete atrioventricular block occurred during the procedure in two patients. In 27 out of 31 attempts (87.1%), new CSP leads were implanted: nine (33.3%) HB pacing leads and 18 (66.7%) LBB area pacing leads.

CONCLUSION

The CSP lead extraction is safe and feasible with a low complication rate and high rate of CSP lead reimplantation.

Для цитирования: Gardas Rafal, Loboda Danuta, Biernat Jolanta, Soral Tomasz, Kulesza Piotr, Gladysz-Wanha Sylwia, Joniec Michal, Sajdok Mateusz, Zub Kamil, Golba Krzysztof S. Extraction of lumenless pacing leads from the His bundle and left bundle branch area: outcomes of the high-volume centre.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2024. DOI: 10.1093/europace/euae213.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.