DOI 10.1016/j.hrthm.2024.06.061

Медицинская библиотека · Heart rhythm · 2024

Safety and performance of conduction system pacing: Real-world experience from a product surveillance registry.

Vijayaraman Pugazhendhi, West Michael, Dresing Thomas, Oren Jess, Abbey Sélim, Zimmerman Patrick, Bauer Rebecca, Butler Kiah, Mangrolia Hardik

СтатьяНаблюдательное исследованиеМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Conduction system pacing (CSP), including His-bundle pacing (HBP) and left bundle branch area pacing (LBBAP), has been used as an alternative for pacemaker indicated patients requiring ventricular pacing.

OBJECTIVE

The purpose of this study was to characterize the safety and performance of HBP and LBBAP among patients enrolled in the Medtronic product surveillance registry (PSR).

METHODS

This observational analysis included patients who underwent pacemaker implantations for HBP or LBBAP with a Medtronic Model 3830 lead between January 2019 and December 2023 in the Medtronic PSR. The primary outcomes were lead-related complications and pacing capture threshold. Baseline characteristics, R-wave amplitude, impedance, and all-cause mortality were summarized.

RESULTS

A total of 2342 patients were included across 77 centers (mean age 74 years; 38.9% female). Of the patients implanted with a 3830 lead for CSP, 64.1% (n = 1502) had LBBAP placement and 35.9% (n = 840) had HBP placement. The most commonly reported indications for CSP were sinus node dysfunction (67.0%) and atrioventricular block (57.2%). LBBAP had lower pacing thresholds, higher R-wave sensing, and higher impedance (all P <.001) through 30 months. At 36 months postimplant, the lead complication rate for LBBAP and HBP was 2.5% and 6.3%, ,respectively with no difference in all-cause mortality.

CONCLUSION

In a multicenter cohort of LBBAP and HBP patients treated with the catheter-delivered 3830 lead, lead-related complication rates were low and electrical parameters were stable through 30 months.

Для цитирования: Vijayaraman Pugazhendhi, West Michael, Dresing Thomas, Oren Jess, Abbey Sélim, Zimmerman Patrick, Bauer Rebecca, Butler Kiah, Mangrolia Hardik. Safety and performance of conduction system pacing: Real-world experience from a product surveillance registry.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.06.061.