Palmisano Pietro, Dell'Era Gabriele, Guerra Federico, Ammendola Ernesto, Ziacchi Matteo, Laffi Mattia, Donateo Paolo, Guido Alessandro, Ghiglieno Chiara, Parlavecchio Antonio, Dello Russo Antonio, Nigro Gerardo, Biffi Mauro, Gaggioli Germano, Senes Jacopo, Patti Giuseppe, Accogli Michele, Coluccia Giovanni

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac resynchronization therapy (CRT) with biventricular pacing (BVP) is a well-established therapy in patients with reduced left ventricular ejection fraction, heart failure, and left bundle branch block. Left bundle branch area pacing (LBBAP) has recently been shown to be a feasible and effective alternative to BVP. Comparative data on the risk of complications between LBBAP and BVP among patients undergoing CRT are lacking.

OBJECTIVE

The aim of this study was to compare the long-term risk of procedure-related complications between LBBAP and BVP in a cohort of patients undergoing CRT.

METHODS

This prospective, multicenter, observational study enrolled 668 consecutive patients (mean age 71.2 ± 10.0 years; 52.2% male; 59.4% with New York Heart Association class III-IV heart failure symptoms) with left ventricular ejection fraction 33.4% ± 4.3% who underwent BVP (n = 561) or LBBAP (n = 107) for a class I or II indication for CRT. Propensity score matching for baseline characteristics yielded 93 matched pairs. The rate and nature of intraprocedural and long-term post-procedural complications occurring during follow-up were prospectively collected and compared between the 2 groups.

RESULTS

During a mean follow-up of 18 months, procedure-related complications were observed in 16 patients: 12 in BVP (12.9%) and 4 in LBBAP (4.3%) (P = .036). Compared with patients who underwent LBBAP, those who underwent BVP showed a lower complication-free survival (P = .032). In multivariate analysis, BVP resulted an independent predictive factor associated with a higher risk of complications (hazard ratio 3.234; P = .042). Complications related to the coronary sinus lead were most frequently observed in patients who underwent BVP (50.0% of all complications).

CONCLUSION

LBBAP was associated with a lower long-term risk of device-related complications compared with BVP in patients with an indication for CRT.

Для цитирования: Palmisano Pietro, Dell'Era Gabriele, Guerra Federico, Ammendola Ernesto, Ziacchi Matteo, Laffi Mattia, Donateo Paolo, Guido Alessandro, Ghiglieno Chiara, Parlavecchio Antonio, Dello Russo Antonio, Nigro Gerardo, Biffi Mauro, Gaggioli Germano, Senes Jacopo, Patti Giuseppe, Accogli Michele, Coluccia Giovanni. Complications of left bundle branch area pacing compared with biventricular pacing in candidates for resynchronization therapy: Results of a propensity score-matched analysis from a multicenter registry.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.02.053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.