Short-term outcomes of micropulse transscleral laser therapy using the revised delivery probe in refractory glaucoma.

Akiyama Takuya, Fujishiro Takashi, Sugimoto Koichiro, Sakata Rei, Saito Hitomi, Honjo Megumi, Aihara Makoto

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the effectiveness of MicroPulse® transscleral laser therapy (MP-TLT) using both the original and revised MicroPulse P3® delivery probes in patients with refractory glaucoma.

STUDY

DESIGN

Retrospective study

METHODS

We analyzed the data of patients with refractory glaucoma who underwent MP-TLT with the original and with the revised MicroPulse P3® Delivery Device from January 2018 to December 2020 at the University of Tokyo Hospital and who were followed up for 3 months. Patients' demographics, preoperative and postoperative intraocular pressure (IOP), medication scores, best-corrected visual acuity (BCVA), and complications were analyzed.

RESULTS

This study enrolled 40 patients in both probe groups. At baseline, the mean IOP was 31.1 mmHg in the original group and 29.2 mmHg in the revised group, and it decreased to 22.6 mmHg and 21.6 mmHg after 3 months, respectively (paired t-test p=0.001, 0.001). No significant difference was noted between the number of medications and BCVA at baseline and 3 months in either probe group (paired t-test, p>0.05). Nor were any serious complications observed.

CONCLUSION

MP-TLT using the revised MicroPulse P3® Delivery Device for the treatment of refractory glaucoma was effective in reducing IOP across glaucoma types with an excellent safety profile and was well-tolerated by the patients.

Для цитирования: Akiyama Takuya, Fujishiro Takashi, Sugimoto Koichiro, Sakata Rei, Saito Hitomi, Honjo Megumi, Aihara Makoto. Short-term outcomes of micropulse transscleral laser therapy using the revised delivery probe in refractory glaucoma.. Japanese journal of ophthalmology. 2022. DOI: 10.1007/s10384-022-00938-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.