EYEMATE-SC Trial: Twelve-Month Safety, Performance, and Accuracy of a Suprachoroidal Sensor for Telemetric Measurement of Intraocular Pressure.

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Szurman Peter, Gillmann Kevin, Seuthe Anna-Maria, Dick H Burkhard, Hoffmann Esther M, Mermoud Andre, Mackert Marc J, Weinreb Robert N, Rao Harsha L, Mansouri Kaweh, EYEMATE-SC Study Group

Аннотация

PURPOSE: Measuring and controlling intraocular pressure (IOP) provide the foundation for glaucoma treatment. Self-tonometry has been proposed as an alternative to measure IOP throughout the entire day better. The novel EYEMATE-SC sensor (Implandata) is implanted in the suprachoroidal space to enable contactless continual IOP monitoring. The aim of the present study was to investigate the 1-year safety, performance, and accuracy of the EYEMATE-SC in patients with primary open-angle glaucoma undergoing simultaneous nonpenetrating glaucoma surgery (NPGS).

DESIGN: Prospective, multicenter, open-label, single-arm, interventional clinical trial.

PARTICIPANTS: Twenty-four eyes of 24 patients with primary open-angle glaucoma who were due to undergo NPGS (canaloplasty or deep sclerectomy).

METHODS: An EYEMATE-SC sensor was implanted during NPGS. Goldmann applanation tonometry (GAT) measurements were compared with the sensors' IOP measurements at all postoperative visits through 12 months.

MAIN OUTCOME MEASURES: Device position and adverse events.

RESULTS: Fifteen eyes underwent canaloplasty, and 9 underwent deep sclerectomy. Successful implantation of the sensor was achieved in all eyes with no reported intraoperative difficulties. Through the 12-month follow-up, no device migration, dislocation, or serious device-related complications were recorded. A total of 536 EYEMATE-SC measurements were pairwise included in the IOP agreement analysis. The overall mean difference between GAT and EYEMATE-SC measurements was 0.8 mmHg (95% confidence interval [CI] of the limits of agreement [LoA], -5.1 to 6.7 mmHg). The agreement gradually improved, and from 3 months after surgery until the end of the follow-up, the mean difference was -0.2 mmHg (95% CI of LoA, -4.6 to 4.2 mmHg) over a total of 264 EYEMATE-SC measurements, and 100% of measurements were within ±5 mmHg of GAT.

CONCLUSIONS: The EYEMATE-SC sensor was safe and well tolerated through 12 months. Moreover, it allowed accurate, continuous IOP monitoring.

FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references.