Comparison of saddle-shape flexibility and elliptical-shape stability between Cosgrove-Edwards and Memo-3D annuloplasty rings using three-dimensional analysis software.

Tsuneto Akira, Eishi Kiyoyuki, Miura Takashi, Tanigawa Kazuyoshi, Matsukuma Seiji, Minami Takako, Koide Yuji, Ikeda Satoshi, Kawano Hiroaki, Maemura Koji

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare three-dimensional dynamics between implanted Cosgrove-Edwards and Sorin Memo-3D annuloplasty rings during the cardiac cycle.

METHODS

We examined 11 Cosgrove-Edwards rings and 20 Sorin Memo-3D rings after mitral plasty using real-time three-dimensional transesophageal echocardiography. We evaluated ring height, ellipticity, and geometry during one cardiac cycle. Four evenly spaced phases each selected during systole and diastole were assessed using REAL VIEW software.

RESULTS

The height of the Cosgrove-Edwards and Sorin Memo-3D rings was similar (2.3 ± 0.8 vs. 1.9 ± 0.9 mm, p = 0.44). The maximum difference in ring height during one cardiac cycle (change in height) was larger for the Cosgrove-Edwards than the Sorin Memo-3D rings (2.3 ± 0.8 vs. 1.5 ± 0.6 mm, p = 0.014). Ellipticity and the maximum difference in ellipticity during one cardiac cycle (change in ellipticity) were larger for Cosgrove-Edwards than Sorin Memo-3D rings (80.0 ± 9.1 vs. 72.0 ± 4.8 %, p = 0.014, respectively, and 12.0 ± 3.1 vs. 6.0 ± 1.8 %, p < 0.001).

CONCLUSIONS

Cosgrove-Edwards rings were more flexible, whereas Sorin Memo-3D rings maintained the elliptical shape more effectively.

Другие публикации по теме

DiastoleEchocardiography, TransesophagealEquipment DesignMitral Valve Annuloplasty
Для цитирования: Tsuneto Akira, Eishi Kiyoyuki, Miura Takashi, Tanigawa Kazuyoshi, Matsukuma Seiji, Minami Takako, Koide Yuji, Ikeda Satoshi, Kawano Hiroaki, Maemura Koji. Comparison of saddle-shape flexibility and elliptical-shape stability between Cosgrove-Edwards and Memo-3D annuloplasty rings using three-dimensional analysis software.. General thoracic and cardiovascular surgery. 2016. DOI: 10.1007/s11748-016-0645-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.