Intraoperative Real-Time Localization of Normal Parathyroid Glands With Autofluorescence Imaging.

Kim Sung Won, Song Seo Hyun, Lee Hyoung Shin, Noh Woong Jae, Oak Chulho, Ahn Yeh-Chan, Lee Kang Dae

СтатьяПеревод пока не подготовлен

Аннотация

CONTEXT: This biomedical investigation is valuable for identification and localization of parathyroid glands during thyroidectomy, which can provide an intraoperative real-time visual guidance.

OBJECTIVE

The objective of the study was to investigate the feasibility of real-time autofluorescence imaging of the parathyroid glands without exogenous contrast dye for their localization and demonstration of relation to the background tissues.

SETTING

This research was undertaken at Kosin University Gospel Hospital.

METHODS

Sixteen normal parathyroid glands from eight patients with papillary thyroid carcinoma were enrolled. Photo images of the surgical field including the parathyroid and background tissues were taken with a digital camera, 780 nm light-emitting diode to excite the parathyroid, and infrared illuminator to visualize the entire neck. The area-averaged autofluorescence intensity of parathyroid over the area-averaged fluorescence intensity of background tissues was measured.

MAIN OUTCOME MEASURE: The location of the parathyroid gland was verified with a single image.

RESULTS

The area-averaged autofluorescence intensity of parathyroid over the area-averaged fluorescence intensity of background tissues for all parathyroid glands was higher than 1, with a minimum of 1.95 and a maximum of 5.20 (average 2.76, SD 0.79). By our technique, all 16 parathyroid glands were detected (positive predictive value of 100%), and the entire surgical field including the parathyroid and background tissues was visualized as well. The parathyroid glands that were exposed or even covered by connective tissues or blood vessels could be detected with strong emission.

CONCLUSIONS

This method showed the precise localization of the parathyroid glands and demonstrated their relation to background tissue. We believe that this simple, nonexogenous dye technique of anatomical guidance can aid surgeons to preserve parathyroid glands during thyroidectomy.

Для цитирования: Kim Sung Won, Song Seo Hyun, Lee Hyoung Shin, Noh Woong Jae, Oak Chulho, Ahn Yeh-Chan, Lee Kang Dae. Intraoperative Real-Time Localization of Normal Parathyroid Glands With Autofluorescence Imaging.. The Journal of clinical endocrinology and metabolism. 2016. DOI: 10.1210/jc.2016-2558.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.