DOI 10.4103/ijo.ijo_2092_22

Медицинская библиотека · Indian journal of ophthalmology · 2023

Comparative evaluation of intraoperative aberrometry and Barrett's toric calculator in toric intraocular lens implantation.

Mukhija Ritika, Vanathi Murugesan, Verma Meena, Raj Nimmy, Gupta Noopur, Tandon Radhika

Наблюдательное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Barrett toric calculator (BTC) is known for its accuracy in toric IOL (tIOL) calculation over standard calculators; however, there is no study in literature to compare it with real-time intraoperative aberrometry (IA). The aim was to compare the accuracy of BTC and IA in predicting refractive outcomes in tIOL implantation.

METHODS

This was an institution-based prospective, observational study. Patients undergoing routine phacoemulsification with tIOL implantation were enrolled. Biometry was obtained from Lenstar-LS 900 and IOL power calculated using online BTC; however, IOL was implanted as per IA (Optiwave Refractive Analysis, ORA, Alcon) recommendation. Postoperative refractive astigmatism (RA) and spherical equivalent (SE) were recorded at one month, and respective prediction errors (PEs) were calculated using predicted refractive outcomes for both methods. The primary outcome measure was a comparison between mean PE with IA and BTC, and secondary outcome measures were uncorrected distance visual acuity (UCDVA), postoperative RA, and SE at one month. SPSS Version-21 was used; P < 0.05 considered significant.

RESULTS

Thirty eyes of 29 patients were included. Mean arithmetic and mean absolute PEs for RA were comparable between BTC (-0.70 ± 0.35D; 0.70 ± 0.34D) and IA (0.77 ± 0.32D; 0.80 ± 0.39D) (P = 0.09 and 0.09, respectively). Mean arithmetic PE for residual SE was significantly lower for BTC (-0.14 ± 0.32D) than IA (0.001 ± 0.33D) (-0.14 ± 0.32D; P = 0.002); however, there was no difference between respective mean absolute PEs (0.27 ± 0.21 D; 0.27 ± 0.18; P = 0.80). At one-month, mean UCDVA, RA, and SE were 0.09 ± 0.10D, -0.57 ± 0.26D, and -0.18 ± 0.27D, respectively.

CONCLUSION

Both IA and BTC give reliable and comparable refractive results for tIOL implantation.

Для цитирования: Mukhija Ritika, Vanathi Murugesan, Verma Meena, Raj Nimmy, Gupta Noopur, Tandon Radhika. Comparative evaluation of intraoperative aberrometry and Barrett's toric calculator in toric intraocular lens implantation.. Indian journal of ophthalmology. 2023. DOI: 10.4103/ijo.ijo_2092_22.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.