Chassain C, Hallak M K, Lesaffre M
Аннотация
PURPOSE: To investigate rotational stability and visual outcomes of patients unilaterally or bilaterally implanted with a new monofocal toric intraocular lens (IOL).
SETTING: Ophthalmology service, clinique Beausoleil, avenue de Lodève, Montpellier.
DESIGN: Single-center retrospective study.
METHODS: This study included patients who underwent routine cataract surgery with the PODEYE toric (BVI/PhysIOL SA, Liège, Belgium) IOL using the ZEISS CALLISTO eye®. Biometry and keratometry data, refractive outcomes, rotational stability, and astigmatism correction were recorded. IOL rotation was evaluated using an image analysis technique. Postoperative assessments were performed at 1 week, 1 month, and 4 to 6 months after surgery.
RESULTS: Clinical outcomes of 102 patients (136 eyes) were analyzed. Patients had a mean age of 74 years. Of the included eyes, 25% had an axial length greater than 24.5mm. Median postoperative IOL rotation from baseline (surgery) was 2̊. With the exception of one outlier (15̊ rotation), IOL rotation was ≤ 6̊ (1 month) and ≤ 10̊ (4-6 months) in 100% of the eyes. No surgical IOL re-positioning was required. Median postoperative corrected distance visual acuity was -0.08 logMAR, and median postoperative subjective cylinder was between 0.25 and 0.50 D.
CONCLUSION: The PODEYE toric IOL showed high rotational stability, allowing for correction of corneal astigmatism during cataract surgery.
SETTING: Ophthalmology service, clinique Beausoleil, avenue de Lodève, Montpellier.
DESIGN: Single-center retrospective study.
METHODS: This study included patients who underwent routine cataract surgery with the PODEYE toric (BVI/PhysIOL SA, Liège, Belgium) IOL using the ZEISS CALLISTO eye®. Biometry and keratometry data, refractive outcomes, rotational stability, and astigmatism correction were recorded. IOL rotation was evaluated using an image analysis technique. Postoperative assessments were performed at 1 week, 1 month, and 4 to 6 months after surgery.
RESULTS: Clinical outcomes of 102 patients (136 eyes) were analyzed. Patients had a mean age of 74 years. Of the included eyes, 25% had an axial length greater than 24.5mm. Median postoperative IOL rotation from baseline (surgery) was 2̊. With the exception of one outlier (15̊ rotation), IOL rotation was ≤ 6̊ (1 month) and ≤ 10̊ (4-6 months) in 100% of the eyes. No surgical IOL re-positioning was required. Median postoperative corrected distance visual acuity was -0.08 logMAR, and median postoperative subjective cylinder was between 0.25 and 0.50 D.
CONCLUSION: The PODEYE toric IOL showed high rotational stability, allowing for correction of corneal astigmatism during cataract surgery.