Rotational stability and clinical outcomes after implantation of a new monofocal toric intraocular lens with double C-loop design.

Chassain C, Hallak M K, Lesaffre M

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate rotational stability and visual outcomes of patients unilaterally or bilaterally implanted with a new monofocal toric intraocular lens (IOL).

SETTING

Ophthalmology service, clinique Beausoleil, avenue de Lodève, Montpellier.

DESIGN

Single-center retrospective study.

METHODS

This study included patients who underwent routine cataract surgery with the PODEYE toric (BVI/PhysIOL SA, Liège, Belgium) IOL using the ZEISS CALLISTO eye®. Biometry and keratometry data, refractive outcomes, rotational stability, and astigmatism correction were recorded. IOL rotation was evaluated using an image analysis technique. Postoperative assessments were performed at 1 week, 1 month, and 4 to 6 months after surgery.

RESULTS

Clinical outcomes of 102 patients (136 eyes) were analyzed. Patients had a mean age of 74 years. Of the included eyes, 25% had an axial length greater than 24.5mm. Median postoperative IOL rotation from baseline (surgery) was 2̊. With the exception of one outlier (15̊ rotation), IOL rotation was ≤ 6̊ (1 month) and ≤ 10̊ (4-6 months) in 100% of the eyes. No surgical IOL re-positioning was required. Median postoperative corrected distance visual acuity was -0.08 logMAR, and median postoperative subjective cylinder was between 0.25 and 0.50 D.

CONCLUSION

The PODEYE toric IOL showed high rotational stability, allowing for correction of corneal astigmatism during cataract surgery.

Для цитирования: Chassain C, Hallak M K, Lesaffre M. Rotational stability and clinical outcomes after implantation of a new monofocal toric intraocular lens with double C-loop design.. Journal francais d'ophtalmologie. 2023. DOI: 10.1016/j.jfo.2022.07.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.