Vecto-keratometry: determination of anterior corneal astigmatism in manual keratometers using power vectors.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Salvador-Roger Raquel, Vila-Andrés Rosa, Micó Vicente, Esteve-Taboada José J
Аннотация
BACKGROUND: A new keratometric routine that employs power vector management for manual keratometers is described. This study evaluates the agreement of the new proposed keratometric technique with the classical one.
RESEARCH DESIGN AND METHODS: The applicability of a new keratometric routine was verified using Helmholtz's and Javal's keratometers. Results were obtained by two different and well-trained examiners over two different samples, one including 65 and the other 74 eyes, respectively. Both conventional keratometry and the newly proposed routine (named vecto-keratometry) were used in each eye to obtain the results. The clinical agreement between the methods was evaluated using Bland-Altman and Passing-Bablok analysis.
RESULTS: For Helmholtz's keratometer, Bland-Altman plots showed good agreement between methods for both astigmatic components being J0 = 0.04 ± 0.20 D and J45 = -0.07 ± 0.17 D. For Javal's keratometer, Passing-Bablok regression test determined regression line for J0 difference as y0 = 1.03, confidence interval: [0.98, 1.10] and regression line for J45 difference as y45 = 0.97, confidence interval: [0.83, 1.12].
CONCLUSIONS: Vecto-keratometry provides accurate clinical results. It has been demonstrated that there are no significant differences between methods in any of the power vector astigmatic components; thus, both methods can be applied interchangeably.
RESEARCH DESIGN AND METHODS: The applicability of a new keratometric routine was verified using Helmholtz's and Javal's keratometers. Results were obtained by two different and well-trained examiners over two different samples, one including 65 and the other 74 eyes, respectively. Both conventional keratometry and the newly proposed routine (named vecto-keratometry) were used in each eye to obtain the results. The clinical agreement between the methods was evaluated using Bland-Altman and Passing-Bablok analysis.
RESULTS: For Helmholtz's keratometer, Bland-Altman plots showed good agreement between methods for both astigmatic components being J0 = 0.04 ± 0.20 D and J45 = -0.07 ± 0.17 D. For Javal's keratometer, Passing-Bablok regression test determined regression line for J0 difference as y0 = 1.03, confidence interval: [0.98, 1.10] and regression line for J45 difference as y45 = 0.97, confidence interval: [0.83, 1.12].
CONCLUSIONS: Vecto-keratometry provides accurate clinical results. It has been demonstrated that there are no significant differences between methods in any of the power vector astigmatic components; thus, both methods can be applied interchangeably.