DOI 10.1097/iae.0000000000003799
Scleral Suture Intraocular Lens Fixation Under Microscope-Intraoperative Optical Coherence Tomography.
Аннотация
PURPOSE: The aim of this study was to evaluate the accessibility and potential value of intraoperative optical coherence tomography (iOCT) during scleral suture intraocular lens (IOL) fixation.
METHODS
This was a prospective cohort study in the Department of Ophthalmology, Eye and ENT Hospital, Fudan University, China. Seven eyes with insufficient capsular support and undergoing two-point scleral suture IOL fixation were included. The potential value of iOCT was evaluated, as well as the safety and efficacy of the surgery.
RESULTS
Seven eyes were included. With a tailor-made iOCT, the structure of the anterior segment could be clearly visualized during the surgery. Intraoperatively, iOCT helped locate the proper place for fixation and access the position of the IOL. After an average 4.43-month follow-up, the spherical equivalent changed significantly ( P < 0.001), but the intraocular pressure, best-corrected visual acuity, and endothelial cell density remained unchanged ( P > 0.05). The IOL was well-centered with a horizontal and a vertical tilt of 0.74 ± 0.60° and 1.13 ± 0.65°, and decentration of 0.28 ± 0.12 mm and 0.30 ± 0.13 mm. The estimated IOL-induced astigmatism was -0.11 diopters (D) ± 0.46 D.
CONCLUSION
Real-time high-resolution images of the anterior segment acquired by the iOCT helped the surgeon to achieve satisfactory results in scleral suture IOL fixation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.