Measurements of Anterior and Posterior Corneal Curvatures with OCT and Scheimpflug Biometers in Patients with Low Total Corneal Astigmatism.
Аннотация
Background
Posterior keratometry measurements are evolving features of the optical biometers. The differences between devices have bigger impact for the low astigmatism values. The majority of adults present the corneal astigmatism below 1.5 D.
Objectives
To compare the total corneal astigmatism measured with two different technologies in cataract patients with corneal astigmatism below 1.5 D. Material and
Methods
Three automated exams were performed on each of the two devices: swept-source optical coherence tomography (SS-OCT) and Scheimpflug biometers. The anterior and total corneal astigmatism and power were analysed. Statistical comparisons were performed for within-subject standard deviation, repeatability, Bland−Altman and vector analysis.
Results
Twenty-nine eyes of twenty-seven patients were included. The limits of agreement between anterior and total corneal astigmatism were narrower for the SS-OCT than for the Scheimpflug biometer (−0.16 to 0.29 D and −0.40 to 0.39 D, respectively). The >0.5 D difference between SS-OCT and Scheimpflug total astigmatism was noticed in 5 (17%) of cases. The difference between mean total keratometric power for both devices was statistically significant (0.2 D, p < 0.001). SS-OCT total corneal flat measurements had worse repeatability than Scheimpflug (p = 0.007).
Conclusions
For the corneal astigmatism <1.5 D, the difference between anterior and total corneal astigmatism measured with SS-OCT was clinically not significant. The mean anterior and total keratometry values obtained with Scheimpflug and SS-OCT biometers are not interchangeable.