DOI 10.1080/00365513.2026.2671961

Медицинская библиотека · Scandinavian journal of clinical and laboratory investigation · 2026

Hemoglobin fraction quantitation with Capillarys 2 FP HbA1c mode in patients with known sickle cell disease.

Klingenberg Olav, Fjeld Bente, Lundbye Ingunn B O

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the feasibility of using Sebia Capillarys HbA1c mode for quantifying hemoglobin fractions in patients with known sickle cell disease to reduce turnaround time.

METHODS

Hemoglobin fractions were analyzed in 58 samples from patients with previously diagnosed sickle cell disease using the Sebia Capillarys 2 FP HbA1c mode (A1c mode) and Sebia Capillarys 2 FP, HEMOGLOBIN(E) mode (Hb mode). Method comparison was performed using Bland-Altman and linear regression analyses. Precision was determined by repeated analysis of internal control material and patient samples, while linearity was studied by serial dilution of a sample containing HbS with a sample without HbS.

RESULTS

The A1c mode showed good agreement with the Hb mode for quantification of HbA, HbS, and HbF fractions, with small biases of 1.44%, -1.66%, and 0.-02% respectively. Good linearity was demonstrated for HbS and HbA quantification (R2=0.9995).

CONCLUSION

The A1c mode is appropriate for quick measurement of hemoglobin fractions in patients with known sickle cell disease, allowing for improved turnaround times in clinical settings. Due to occasional peak shifts and peak mislabeling by the software, a flow chart is included to guide proper A1c mode electropherogram interpretation.

Для цитирования: Klingenberg Olav, Fjeld Bente, Lundbye Ingunn B O. Hemoglobin fraction quantitation with Capillarys 2 FP HbA1c mode in patients with known sickle cell disease.. Scandinavian journal of clinical and laboratory investigation. 2026. DOI: 10.1080/00365513.2026.2671961.