DOI 10.1097/00002341-199903000-00002
Exposure of expanded polytetrafluoroethylene-wrapped hydroxyapatite orbital implant: a report of two patients.
Аннотация
PURPOSE: Hydroxyapatite (HA) spheres used to replace volume after an enucleation are often wrapped with autologous tissue before orbital implantation. Man-made materials are less expensive and pose no risk for viral transmission. The use of expanded polytetrafluoroethylene (ePTFE) to wrap HA spheres was evaluated.
METHODS
The medical records of 2 consecutive patients who underwent uncomplicated implantation of an HA sphere wrapped in ePTFE were reviewed.
RESULTS
An unusual reaction to the ePTFE material that was nonresponsive to topical or systemic antibiotic therapy developed in these 2 patients. Eventual wound erosion and bacterial infection of the implant necessitated its removal.
CONCLUSIONS
Although well tolerated in other surgeries, ePTFE, when used to wrap HA spheres and placed into the orbit, may cause persistent conjunctival discharge, pyogenic granuloma formation, and eventual wound erosion. Therefore, the use of this material to wrap HA spheres is not recommended.