Development of a laboratorial platform for diagnosis of schistosomiasis mansoni by PCR-ELISA.

Senra Carolina, Gomes Luciana Inácia, Siqueira Liliane Maria Vidal, Coelho Paulo Marcos Zech, Rabello Ana, Oliveira Edward

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

We developed a laboratorial platform to release a commercial platform used in the PCR-ELISA for the molecular diagnosis of schistosomiasis mansoni. On following, PCR-ELISA platform laboratorial was evaluated in 206 feces samples collected of individual living in a Brazilian low endemicity area.

RESULTS

The PCR-ELISA laboratorial platform indicated a prevalence rate of 25.2%, which was higher than the Kato-Katz technique (18.4%) and lower than the commercial platform (30.1%). Considering Kato-Katz technique as the reference, there were 97.4% and 91.1% of relative sensitivity and specificity rates, respectively. The laboratorial platform presented good precision, performance diagnostic, and can be used in replacement to the commercial platform for diagnosis of schistosomiasis by PCR-ELISA.

Для цитирования: Senra Carolina, Gomes Luciana Inácia, Siqueira Liliane Maria Vidal, Coelho Paulo Marcos Zech, Rabello Ana, Oliveira Edward. Development of a laboratorial platform for diagnosis of schistosomiasis mansoni by PCR-ELISA.. BMC research notes. 2018. DOI: 10.1186/s13104-018-3571-7.