Sustainable four-dimensional cardiac CT: Minimizing data storage via reduced matrix size with optimized reconstruction.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Nakajima H, Nishii T, Okuyama S, Tanaka T, Namboku T, Hara K, Hayashi T, Morikawa T, Ohta Y, Fukuda T

Аннотация

INTRODUCTION: The increasing volume of four-dimensional cardiac CT data complicates transcatheter aortic valve repair (TAVR) planning, particularly regarding data storage. This study evaluates a modified 256 × 256 reconstruction method for TAVR-CT that reduces storage requirements while maintaining aortic valve measurement accuracy.

METHODS: A retrospective analysis was performed on 75 TAVR-CT scans obtained using the dual-source energy-integrating detector CT. The modified 256 × 256 reconstruction used a sharper Bv44 kernel (iterative construction strength 4) and narrowed the field of view to 80 %. We assessed virtual basal ring (BR) areas using reference 512 × 512, standard 256 × 256, and modified 256 × 256 images. Five technologists-three from the CT team and two from the interventional radiology (IR) team-manually evaluated the BR areas. The measurements were averaged within each team. The intraclass correlation coefficient (ICC), bias, and consistency in device selection were assessed.

RESULTS: Among 75 patients (48 women, median age 85), the mean BR area was 416.2 ± 77.9 mm2. The mean bias between the modified 256 × 256 and reference 512 × 512 images was -2.5 mm2 (95 % CI: -5.2 to 0.3), significantly smaller (P = 0.04) than the bias between standard 256 × 256 and reference (4.1 mm2, 95 % CI: 0.7 to 7.5). BR measurements demonstrated excellent reproducibility (ICC = 0.996-0.997), with limits of agreement (LOA) for the modified 256 × 256 ranging from -13.0 to 12.2 mm2 for the IR team and -14.5 to 11.6 mm2 for the CT team, comparable to the inter-team LOA of -18.1 to 13.1 mm2. Valve selection consistency was 95 % (Cohen's Kappa = 0.92).

CONCLUSION: The modified 256 × 256 reconstruction method effectively preserved BR measurements and valve selection reliability.

IMPLICATIONS FOR PRACTICE: The proposed 256 × 256 method could cut data storage while ensuring accurate valve selection.