Ultra-wide field imaging system and traditional retinal examinations for screening fundus changes after cataract surgery.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Peng Jie, Zhang Qi, Jin Hai-Ying, Lu Wu-Yi, Zhao Pei-Quan
Аннотация
AIM: To compare the results of non-mydriatic ultra-wide field imaging system, mydriatic slit-lamp lens (Volk +90 D) and mydriatic Goldmann three-mirror contact lens examinations in screening fundus lesions among patients after cataract surgery.
METHODS: Non-mydriatic images were obtained with an Optomap panoramic 200Tx (Optomap 200Tx) 3d after surgery and graded by a blinded ophthalmologist. A mydriatic slit-lamp lens examination was performed by another blinded retinal specialist on the same day. A third blinded retinal specialist examined patients two weeks after surgery using a Goldmann three-mirror contact lens.
RESULTS: In total, 160 patients (184 eyes) were examined, and 66, 69, and 75 cases of retinal lesion(s) were identified using the Optomap 200Tx, slit-lamp lens, and Goldmann three-mirror contact lens, respectively. In 13 cases, fundus changes were sight-threatening. The results obtained by Optomap 200Tx examination and by mydriatic slit-lamp lens examination have good consistency (P=0.375, Kappa=0.942). The mydriatic Goldmann three-mirror lens examination revealed more fundus lesions but are consistent with Optomap 200Tx (P=0.004, Kappa=0.897) and mydriatic slit-lamp lens examination (P=0.031, Kappa=0.932).
CONCLUSION: Early post-operative fundus screening in cataract patients is extremely important and necessary to prevent further vision loss. Wide-field imaging is a feasible and convenient tool for fundus examination that can be used as a primary screening method among patients after cataract surgery.
METHODS: Non-mydriatic images were obtained with an Optomap panoramic 200Tx (Optomap 200Tx) 3d after surgery and graded by a blinded ophthalmologist. A mydriatic slit-lamp lens examination was performed by another blinded retinal specialist on the same day. A third blinded retinal specialist examined patients two weeks after surgery using a Goldmann three-mirror contact lens.
RESULTS: In total, 160 patients (184 eyes) were examined, and 66, 69, and 75 cases of retinal lesion(s) were identified using the Optomap 200Tx, slit-lamp lens, and Goldmann three-mirror contact lens, respectively. In 13 cases, fundus changes were sight-threatening. The results obtained by Optomap 200Tx examination and by mydriatic slit-lamp lens examination have good consistency (P=0.375, Kappa=0.942). The mydriatic Goldmann three-mirror lens examination revealed more fundus lesions but are consistent with Optomap 200Tx (P=0.004, Kappa=0.897) and mydriatic slit-lamp lens examination (P=0.031, Kappa=0.932).
CONCLUSION: Early post-operative fundus screening in cataract patients is extremely important and necessary to prevent further vision loss. Wide-field imaging is a feasible and convenient tool for fundus examination that can be used as a primary screening method among patients after cataract surgery.