Assessment of cardiac implantable electric device lead perforation using a metal artifact reduction algorithm in cardiac computed tomography.

Kidoh Masafumi, Oda Seitaro, Nakato Kengo, Sakabe Daisuke, Kanazawa Hisanori, Takashio Seiji, Nakaura Takeshi, Nagayama Yasunori, Sasao Akira, Hatemura Masahiro, Funama Yoshinori, Kaikita Koichi, Tsujita Kenichi, Ikeda Osamu, Azuma Minako, Hirai Toshinori

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: CT is considered the non-invasive gold standard for evaluating cardiac implantable electronic devices (CIEDs) lead perforation, but metal artifacts caused by the lead tip affect the image quality and make a definitive diagnosis challenging. We compared the performances of the metal artifact reduction (MAR) algorithm and the conventional algorithm for identification of the right ventricular (RV) lead tip position in cardiac CT studies of patients with CIEDs.

METHOD: Forty-seven consecutive patients (26 men; age 70.3 ± 15.4 years) with CIEDs underwent cardiac CT. Using the conventional and MAR algorithm, two image reconstructions were performed for each scan. We calculated the artifact index (AI) to assess the quantitative capability of the MAR algorithm for artifact reduction and visually assessed the RV lead tip position on both images as follows: non-perforation, perforation, and equivocal.

RESULTS

The mean AIs were significantly lower with the MAR algorithm than with the conventional algorithm (96.7 ± 40.1 HU vs. 284.6 ± 134.1 HU, P < 0.001). Thirteen (27.7 %) patients were diagnosed as equivocal using the conventional algorithm but were diagnosed with perforation (2 patients) and non-perforation (11 patients) using the MAR algorithm (equivocal rate: 27.7 % vs. 0%, P < 0.001). Using the MAR algorithm, all cases were diagnosed with perforation (6 patients, 12.8 %) or non-perforation (41 patients, 87.2 %).

CONCLUSIONS

The MAR algorithm effectively reduced metal artifacts and allowed us to diagnose the presence or absence of perforation in all cases, whereas definitive diagnosis was difficult with the use of conventional algorithm in 27.7 % of cases.

Для цитирования: Kidoh Masafumi, Oda Seitaro, Nakato Kengo, Sakabe Daisuke, Kanazawa Hisanori, Takashio Seiji, Nakaura Takeshi, Nagayama Yasunori, Sasao Akira, Hatemura Masahiro, Funama Yoshinori, Kaikita Koichi, Tsujita Kenichi, Ikeda Osamu, Azuma Minako, Hirai Toshinori. Assessment of cardiac implantable electric device lead perforation using a metal artifact reduction algorithm in cardiac computed tomography.. European journal of radiology. 2021. DOI: 10.1016/j.ejrad.2021.109530.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.