Clinical biochemistry · 2020

Change in troponin concentrations in patients with macrotroponin: An in vitro mixing study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lam Leo, Heron Campbell, Aspin Lisa, Ha Leah, Kyle Campbell V

Аннотация

INTRODUCTION: Macrotroponin is a complex formed between endogenous cardiac troponin autoantibodies (cTnAABs) and circulating cardiac troponin (cTn). The potential effect of macrotroponin on current high sensitivity cTn assays has not been fully explored but has recently been identified as a major cause of discrepancy in cTn results between assays. In this study we investigated the effects of mixing troponin (cTn) standards to specimens with and without macrotroponin.

METHOD: Macrotroponin was identified in specimens by a recovery of cTnI < 40% following protein A immunoglobulin depletion. Troponin standards containing cTn-IC and cTn-TIC complexes were mixed with serum samples, with (n = 20) and without (n = 10) the presence of macrotroponin. Specimens were tested for cTn before and after mixing by three commercially available high sensitivity cTn assays. Gel filtration chromatography was carried out on five specimens with macrotroponin and each fraction was analzyed by multiple cTn assays.

FINDINGS: Following mixing with cTn-TIC standard, all specimens with macrotroponin had a markedly reduced absolute increase in cTnI, indicating negative analytical interference due to macrotroponin. Following mixing with the cTn-IC standard, specimens with macrotroponin demonstrated highly variable changes in cTnI, suggesting significant heterogeneity in macrotroponin complex reactivity between individuals. When the ratio of change, calculated by dividing the absolute change between two cTn assays, was compared between specimens with and without macrotroponin, significant differences were observed (p < 0.001). These findings were supported by variable migration of peak cTn activity on gel filtration chromatography.

CONCLUSION: Macrotroponin leads to assay dependent analytical interference affecting current high sensitivity troponin I assays. Furthermore, endogenously occurring cTnAABs are conformationally specific and the analytical effects vary between assays and individuals.