MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrtlng.2020.06.016

Медицинская библиотека · Heart & lung : the journal of critical care · 2020

Bubbles-in-the-chamber vs digital screen in chest drainage: A blind analysis of compared postoperative air leaks evaluation.

Marulli Giuseppe, Brascia Debora, De Iaco Giulia, Comacchio Giovanni Maria, Natale Giuseppe, Nosotti Mario, Mendogni Paolo, Pieropan Sara, Lopez Camillo, Di Rienzo Gaetano, Andriolo Luigi Gaetano, Rea Federico

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Chest drainage systems are affected by intra and inter-observer variability and poor sensibility in detecting minimal or apparent air leaks.

OBJECTIVES

Overcome intra and inter-observer variability in detecting air leaks.

METHODS

After surgery, a single apical chest tube was connected to the Drentech™ PalmEVO device and air leaks were checked twice a day by observation of both bubbles-in-the-chamber and digital data.

RESULTS

On a total of 624 observations, disagreement between digital and traditional systems was recorded in 60(9.6%) cases. In 25(21.4%) patients, a disagreement was recorded. Overall, the digital evaluation influenced clinical management in 13(52%). In 10(40%) patients with temporary discordant features, the presence of high pleural fluid output led to a progressive final concordance.

CONCLUSIONS

Disagreement between traditional and digital systems in checking air leaks is not negligible. Digital systems could give advantages in making an objective assessment of air leaks, standardizing the timing of chest tube removal.

Для цитирования: Marulli Giuseppe, Brascia Debora, De Iaco Giulia, Comacchio Giovanni Maria, Natale Giuseppe, Nosotti Mario, Mendogni Paolo, Pieropan Sara, Lopez Camillo, Di Rienzo Gaetano, Andriolo Luigi Gaetano, Rea Federico. Bubbles-in-the-chamber vs digital screen in chest drainage: A blind analysis of compared postoperative air leaks evaluation.. Heart & lung : the journal of critical care. 2020. DOI: 10.1016/j.hrtlng.2020.06.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.