m+Medelevator.RU
← Поиск публикаций
JACC. Clinical electrophysiology · 2020

Outcomes of Standard Permanent Active Fixation Leads for Temporary Pacing.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Cipriano Ralph, Gupta Ashwani, Subzposh Faiz, McCaffrey James A, Koman Eduard, Fridman David, Saltzman Heath, Kutalek Steven P

Аннотация

OBJECTIVES: This study investigated the performance of Temporary Pacing via an Externalized Active-Fixation (TPEAF) lead.

BACKGROUND: The incidence of cardiac implantable electronic device infections is increasing, which necessitates the need for transvenous lead extraction (TLE). Pacemaker-dependent patients require temporary pacing during the guideline-recommended waiting period before reimplantation. Data regarding safety and efficacy of TPEAF leads are very limited.

METHODS: We evaluated patients implanted with TPEAF leads post-TLE at our center between April 2004 and December 2017.

RESULTS: TPEAF leads were placed in 158 patients. The mean age was 74 ± 11 years. The median duration of the temporary lead was 6 days (range 1 to 29). There were 4 procedural complications (2.5% incidence): 1 patient had cardiac arrest from hyperkalemia, 2 developed cardiac tamponade, and 1 had profuse bleeding from the entry point of the leads. There were 13 complications post-implantation (8.2% incidence): 8 lead dislodgments, 1 elevated pacing threshold, 2 vegetations on the temporary lead, 1 pneumothorax, and 1 loss of capture due to the generator "safety switch." All dislodgements occurred within 24 h, except 1 on day 3. Sixteen patients died during the hospital stay: 10 due to septic shock, 2 due to hyperkalemic cardiac arrest, 3 due to ventricular tachycardia, and 1 due to a massive cerebrovascular accident.

CONCLUSIONS: The use of TPEAF leads is safe and efficacious in pacemaker-dependent patients post-TLE. Dislodgement can occur within the first 24 h. The presence of persistent fever and positive blood cultures should raise concern for vegetation on the temporary lead.