DOI 10.1016/j.resuscitation.2018.09.024

Медицинская библиотека · Resuscitation · 2018

Enhancement of capnogram waveform in the presence of chest compression artefact during cardiopulmonary resuscitation.

Ruiz de Gauna Sofía, Leturiondo Mikel, Gutiérrez J Julio, Ruiz Jesus M, González-Otero Digna M, Russell James K, Daya Mohamud

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Current resuscitation guidelines emphasize the use of waveform capnography to help guide rescuers during cardiopulmonary resuscitation (CPR). However, chest compressions often cause oscillations in the capnogram, impeding its reliable interpretation, either visual or automated. The aim of the study was to design an algorithm to enhance waveform capnography by suppressing the chest compression artefact.

METHODS

Monitor-defibrillator recordings from 202 patients in out-of-hospital cardiac arrest were analysed. Capnograms were classified according to the morphology of the artefact. Ventilations were annotated using the transthoracic impedance signal acquired through defibrillation pads. The suppression algorithm is designed to operate in real-time, locating distorted intervals and restoring the envelope of the capnogram. We evaluated the improvement in automated ventilation detection, estimation of ventilation rate, and detection of excessive ventilation rates (over-ventilation) using the capnograms before and after artefact suppression.

RESULTS

A total of 44 267 ventilations were annotated. After artefact suppression, sensitivity (Se) and positive predictive value (PPV) of the ventilation detector increased from 91.9/89.5% to 98.0/97.3% in the distorted episodes (83/202). Improvement was most noticeable for high-amplitude artefact, for which Se/PPV raised from 77.6/73.5% to 97.1/96.1%. Estimation of ventilation rate and detection of over-ventilation also upgraded. The suppression algorithm had minimal impact in non-distorted data.

CONCLUSION

Ventilation detection based on waveform capnography improved after chest compression artefact suppression. Moreover, the algorithm enhances the capnogram tracing, potentially improving its clinical interpretation during CPR. Prospective research in clinical settings is needed to understand the feasibility and utility of the method.

Для цитирования: Ruiz de Gauna Sofía, Leturiondo Mikel, Gutiérrez J Julio, Ruiz Jesus M, González-Otero Digna M, Russell James K, Daya Mohamud. Enhancement of capnogram waveform in the presence of chest compression artefact during cardiopulmonary resuscitation.. Resuscitation. 2018. DOI: 10.1016/j.resuscitation.2018.09.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.