DOI 10.1097/j.jcrs.0000000000001828

Медицинская библиотека · Journal of cataract and refractive surgery · 2026

Comparison of ocular biometry and refractive outcome between a new and classic optical biometer.

Lin Haowen, Xie Xiaohang, Gao Ruoxi, Jing Yashi, Chen Danyue, Zhang Jinhong, Qiu Xiaozhang, Zhang Jiaqing, Luo Lixia

СтатьяСравнительное исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the ocular biometry and refractive outcome between 2 optical biometers.

SETTING

Zhongshan Ophthalmic Center, Guangzhou, China.

DESIGN

Prospective observational study.

METHODS

953 patients with cataract underwent preoperative biometry including ZW-30 sum-of-segments (SOS) method (ZW SM ), ZW-30 composite method (ZW CM ), and IOLMaster 700. Agreement of axial length (AL), with or without Cooke-modified AL (CMAL) adjustment, was analyzed using Bland-Altman 95% limits of agreement (LoA). Subgroup analysis was used based on ALs (short eyes: AL <22 mm; normal eyes: 22 mm ≤ AL <26 mm; long eyes: AL ≥26 mm). Refractive prediction accuracy was evaluated using the Emmetropia verifying optical (EVO) 2.0 formula and its SOS-optimized version (EVO 2.0 SOS ).

RESULTS

In short and normal eyes, narrow 95% LoAs (<0.2 mm) were identified among 3 ALs. However, AL obtained by ZW SM was lower compared with this obtained by ZW CM and IOLMaster 700 (95% LoA -0.39 to 0.01 mm; -0.38 to 0.04 mm) in long eyes. CMAL adjustment enhanced the agreement of AL between ZW SM and ZW CM (95% LoA -0.01 mm to 0.02 mm), ZW SM and IOLMaster 700 (95% LoA -0.10 mm to 0.07 mm) in long eyes. Myopic prediction errors (PE) have been identified in the use of ocular biometric parameters obtained from ZW SM (mean PE [ME]: EVO 2.0, -0.19 diopters [D]; EVO 2.0 SOS , -0.18 D). After adjusting ME to zero, no difference was observed in PE calculated using any combination of formulas based on biometric measurements from 3 devices.

CONCLUSIONS

This new segmented biometer demonstrated excellent agreement with IOLMaster 700 in short and normal eyes. However, ALs obtained by IOLMaster 700 are not interchangeable with the SOS method and require CMAL adjustment in long eyes. The application of the SOS method's ocular biometric parameters in refractive prediction led to myopic errors, which suggests constant optimization.

Для цитирования: Lin Haowen, Xie Xiaohang, Gao Ruoxi, Jing Yashi, Chen Danyue, Zhang Jinhong, Qiu Xiaozhang, Zhang Jiaqing, Luo Lixia. Comparison of ocular biometry and refractive outcome between a new and classic optical biometer.. Journal of cataract and refractive surgery. 2026. DOI: 10.1097/j.jcrs.0000000000001828.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.