BACKGROUND

De novo aortic insufficiency (AI) is a common adverse event after continuous-flow left ventricular assist device (LVAD) placement and is associated with morbidity and mortality. This study aims to compare the development of de novo AI between HeartMate 3 (Abbott) and HeartMate II LVAD recipients.

METHODS

A retrospective review was conducted of clinical characteristics and serial echocardiograms (1 month, 6 months, and 1 year postimplantation) of HeartMate 3 patients implanted between November 2014 and March 2019 and of HeartMate II patients implanted between April 2004 and December 2015 at Columbia University Irving Medical Center. One hundred twenty-two patients were excluded from analysis for a history of aortic valve surgery, concomitant aortic valve surgery with LVAD implant, or more than trace preoperative AI left untreated. De novo AI was defined as development of more than mild AI after LVAD implant.

RESULTS

Included in the study were 121 HeartMate 3 patients and 270 HeartMate II patients. After accounting for competing risks of death and transplantation, there was no significant difference in the development of de novo AI by 1 year postimplantation between HeartMate II and HeartMate 3 patients (P = .68). There was no significant difference in severity of AI developed up to 1 year post-implantation between HeartMate II and HeartMate 3 patients.

CONCLUSIONS

Development of de novo AI is comparable between HeartMate 3 and HeartMate II patients. There is no significant difference in severity of AI between HeartMate II and HeartMate 3 patients.

Для цитирования: Malick Alyyah, Ning Yuming, Kurlansky Paul A, Melehy Andrew, Yuzefpolskaya Melana, Colombo Paolo C, Sayer Gabriel, Uriel Nir, Naka Yoshifumi, Takeda Koji. Development of De Novo Aortic Insufficiency in Patients With HeartMate 3.. The Annals of thoracic surgery. 2021. DOI: 10.1016/j.athoracsur.2021.08.074.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.