MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/opth.s542284

Медицинская библиотека · Clinical ophthalmology (Auckland, N.Z.) · 2025

Comparison of Post-Operative Outcomes between IOP-20 and IOP-50 in Phacoemulsification with Active Fluidics System: Randomized Single Blinded Trial.

Wavikar Chandrashekhar M, Tanna Mamta N, Wavikar Gargi C, Kale Chandrashekhar B, Setia Maninder Singh

Клинический случайКлиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the central corneal thickness, mitigation events, post-operative vision, and complications in patients who underwent cataract surgery using the CENTURION ® Vision System with the ACTIVE SENTRY ® (Alcon Industries, Inc.) system and were assigned to one of two groups: intraocular pressure (IOP) group 20 or 50.

SETTING

Wavikar Eye Institute, Thane, India.

DESIGN

A prospective comparative two arm, randomized, single blinded, interventional study.

METHODS

A total of 107 eyes were randomized to either the IOP-20 (n=55) or IOP-50 (n=52) group. All patients underwent phacoemulsification with an active fluidics system performed by a single surgeon; intraocular pressure (IOP) was set at either 20 or 50 mmHg.

RESULTS

The mean [SD] baseline CCT in the IOP-20 group was 511.9 [35.7] μm and it was 521.1 [34.8] μm in the IOP 50 group (p=0.18). It had changed to 530.6 [39.8] μm in the IOP-20 group and 539.5 [39.2] μm in the IOP 50 (p=0.24). The mean [SD] ECD at the end of day 90 was 2423.0 [204.6] cells/mm2 in the IOP-20 group and 2476.1 [236.0] cells/mm2 in the IOP-50 group (p=0.22). Approximately 22% of the eyes in the IOP-20 group and 77% of the eyes in the IOP-50 group required active surge mitigation; this difference was statistically significant (p<0.001).

CONCLUSION

We found that the phacoemulsification with an active fluidics system handpiece at 20 mmHg provided a safe and stable environment for cataract surgery.

Для цитирования: Wavikar Chandrashekhar M, Tanna Mamta N, Wavikar Gargi C, Kale Chandrashekhar B, Setia Maninder Singh. Comparison of Post-Operative Outcomes between IOP-20 and IOP-50 in Phacoemulsification with Active Fluidics System: Randomized Single Blinded Trial.. Clinical ophthalmology (Auckland, N.Z.). 2025. DOI: 10.2147/opth.s542284.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.