Ruptures of the device landing zone in patients undergoing transcatheter aortic valve implantation: an analysis of TAVI Karlsruhe (TAVIK) patients.
Аннотация
BACKGROUND
Aortic rupture of the device landing zone is a rare complication of transcatheter aortic valve implantation (TAVI) and it is associated with significant mortality.
METHODS
This study reports on the experience of a single-center in a case series of more than 1,000 implants. We explored patient and procedural characteristics aiming at identifying variables that increase the risk for aortic root rupture.
RESULTS
Among a total of 1,000 TAVI procedures, six patients (0.6 %) had a rupture of the device landing zone. Five of these patients received the balloon-expandable Edwards SAPIEN valve (5/813; 0.62 %) of which four had a supraannular and one a subannular rupture. One patient received the self-expanding Medtronic CoreValve (1/199; 0.5 %; p = n.s. vs. SAPIEN) and had an annular rupture. Factors that were associated with aortic rupture were: (1) the relative size of the valve compared with the aortic annulus and its geometric form; (2) the need for post-dilation of the new valve because of paravalvular leakage; and (3) the location and severity of calcification. We determined, to avoid aortic rupture, caution may be necessary in the presence of the following conditions: (1) flat sinuses of Valsalva and severe calcifications of either the body or the free edge of the aortic cusps (supraannular rupture); (2) an ellipsoid annulus and bulky calcifications on either the base of the cusps or the rim of the annulus (annular rupture); and (3) a narrow left-ventricular outflow tract (LVOT) and bulky calcification of the LVOT (subannular rupture). After considering these precautions, we observed no case of aortic root rupture in the following 600 cases.
CONCLUSION
The data indicate that to reduce the frequency of aortic rupture, a careful pre-procedural planning appears essential to avoid this serious and potentially deleterious complication.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.