Journal of cardiac surgery · 2014

Development of a single endovascular device for aortic valve replacement and ascending aortic repair.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rylski Bartosz, Szeto Wilson Y, Bavaria Joseph E, Branchetti Emanuela, Moser William, Milewski Rita K

Аннотация

BACKGROUND AND AIM OF THE STUDY: Transcatheter aortic valve implantation (TAVI) is contraindicated in the presence of an ascending aortic aneurysm. Our aim was to design a composite endovascular device enabling ascending aortic repair and TAVI.

METHODS: From 2007 to 2013, among 1196 patients with severe aortic stenosis screened for TAVI, 79 nonbicuspid patients had ascending aortic diameter >45 mm. Proximal aortic geometry was assessed in those with computed tomography angiography.

RESULTS: All together, 51 patients (35 males, aged 85 ± 8 years; 19 TAVI, 10 open Wheat procedures, 22 managed conservatively) were included. The required annular diameter for implantation of currently available TAVI prostheses was met in 41% (21/51). Novel prosthetic valves appropriate for annular range up to 30 mm would extend device applicability to 78% (40/51). Proximal and distal diameters of the graft-covering portion ranging between 30 and 46 mm would enable 10% graft oversizing in all but six patients. In 88% (45/51) the required minimum 10 mm distance between aortic valve annulus and coronary artery ostia was found. Mean distance between left and right coronary artery ostia and sinotubular junction was 2.6 ± 1.5 and 3.2 ± 1.7 mm, respectively.

CONCLUSIONS: Novel composite endovascular valved grafts may extend the application of transcatheter techniques to patients denied TAVI due to a concomitant ascending aneurysm. The location of coronary arteries in relation to the sinotubular junction must be addressed in designing these composite valve grafts.