DOI 10.1007/s00330-011-2068-8

Медицинская библиотека · European radiology · 2011

Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients.

Habets Jesse, Symersky Petr, van Herwerden Lex A, de Mol Bas A J M, Spijkerboer Anje M, Mali Willem P Th M, Budde Ricardo P J

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

Multidetector CT (MDCT) has shown potential for prosthetic heart valve (PHV) assessment. We assessed the image quality of different PHV types to determine which valves are suitable for MDCT evaluation.

METHODS

All ECG-gated CTs performed in our institutions since 2003 were reviewed for the presence of PHVs. After reconstruction in 3 specific PHV planes, image quality of the supravalvular, perivalvular, subvalvular and valvular regions was scored on a four-point scale (1 = non-diagnostic, 2 = moderate, 3 = good and 4 = excellent) by two independent observers.

RESULTS

Eighty-four CT examinations (66 cardiac, 18 limited-dose aortic protocols) of 83 patients with a total of 91 PHVs in the aortic (n = 71), mitral (n = 17), pulmonary (n = 1) and tricuspid (n = 2) position were included. CT was performed on a 16-slice (n = 4), 64-slice (n = 28) or 256-slice (n = 52) MDCT system. Median image quality scores for the supra-, peri- and subvalvular regions and valvular detail were (3.5, 3.3, 3.5 and 3.5, respectively) for bileaflet PHV; (3.0, 3.0, 3.5 and 3.0, respectively) for Medtronic Hall PHV; (1.0, 1.0, 1.0 and 1.0, respectively) for Björk-Shiley and Sorin monoleaflet PHV and (3.5, 3.5, 4.0 and 2.0 respectively) for biological PHV.

CONCLUSION

Currently implanted PHVs have good image quality on MDCT and are suitable for MDCT evaluation.

Для цитирования: Habets Jesse, Symersky Petr, van Herwerden Lex A, de Mol Bas A J M, Spijkerboer Anje M, Mali Willem P Th M, Budde Ricardo P J. Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients.. European radiology. 2011. DOI: 10.1007/s00330-011-2068-8.