MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mrm.70097

Медицинская библиотека · Magnetic resonance in medicine · 2025

Scanner-based real-time automated volumetry reporting of the fetus, amniotic fluid, placenta, and umbilical cord for fetal MRI at 0.55T.

Neves Silva Sara, Uus Alena, Waheed Hadi, Bansal Simi, St Clair Kamilah, Norman Wendy, Aviles Verdera Jordina, Cromb Daniel, Woodgate Tomas, van Poppel Milou, Steinweg Johannes K, Matthew Jacqueline, Pushparajah Kuberan, Lloyd David, Kyriakopoulou Vanessa, Siassakos Dimitris, David Anna, Hajnal Joseph V, Story Lisa, Rutherford Mary A, Hutter Jana

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This work aims to enable real-time automated intra-uterine volumetric reporting and fetal weight estimation for fetal MRI, deployed directly on the scanner.

METHODS

A multi-region segmentation nnUNet was trained on 146 images of 73 fetal subjects (coronal and axial orientations) for the parcellation of the fetal head, fetal body, placenta, amniotic fluid, and umbilical cord from whole uterus bSSFP stacks. A reporting tool was then developed to integrate the segmentation outputs into an automated report, providing volumetric measurements, fetal weight estimations, and z-score visualizations. The complete pipeline was subsequently deployed on a 0.55T MRI scanner, enabling real-time inference and fully automated reporting in the duration of the acquisition.

RESULTS

The segmentation pipeline was quantitatively and retrospectively evaluated on 36 stacks of 18 fetal subjects and demonstrated sufficient performance for all labels, with high scores ( > $$ > $$ 0.98) for the fetus, placenta, and amniotic fluid, and 0.91 for the umbilical cord. The prospective evaluation of the scanner deployment step was successfully performed on 50 cases, with the regional volumetric reports available directly on the scanner.

CONCLUSIONS

This work demonstrated the feasibility of multi-regional intra-uterine segmentation, fetal weight estimation, and automated reporting in real-time. This study provides a robust baseline solution for the integration of fully automated scanner-based measurements into fetal MRI reports.

Для цитирования: Neves Silva Sara, Uus Alena, Waheed Hadi, Bansal Simi, St Clair Kamilah, Norman Wendy, Aviles Verdera Jordina, Cromb Daniel, Woodgate Tomas, van Poppel Milou, Steinweg Johannes K, Matthew Jacqueline, Pushparajah Kuberan, Lloyd David, Kyriakopoulou Vanessa, Siassakos Dimitris, David Anna, Hajnal Joseph V, Story Lisa, Rutherford Mary A, Hutter Jana. Scanner-based real-time automated volumetry reporting of the fetus, amniotic fluid, placenta, and umbilical cord for fetal MRI at 0.55T.. Magnetic resonance in medicine. 2025. DOI: 10.1002/mrm.70097.