DOI 10.4244/eij-d-24-00619

Медицинская библиотека · EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2024

Redo-TAVI with the SAPIEN 3 valve in degenerated calcified CoreValve/Evolut explants.

Meier David, Nigade Anish, Lai Althea, Dorman Kyle, Gill Hacina, Javani Shahnaz, Akodad Mariama, Wood David A, Rogers Toby, Puri Rishi, Allen Keith B, Chhatriwalla Adnan K, Reardon Michael J, Tang Gilbert H L, Bapat Vinayak N, Webb John G, Fukuhara Shinichi, Sellers Stephanie L

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Redo-transcatheter aortic valve implantation (TAVI) is the treatment of choice for failed transcatheter aortic valves. Currently, implantation of a SAPIEN 3 (S3) is indicated for redo-TAVI in degenerated CoreValve/Evolut (CV/EV) transcatheter aortic valves (TAVs) but is not well understood.

AIMS: We aimed to evaluate S3 function following implantation in explanted calcified CV/EV TAVs and to assess the impact of CV/EV pathology on redo-TAVI outcomes.

METHODS

Ex vivo hydrodynamic testing was performed per the International Organization for Standardization (ISO) 5840-3 standard on 4 S3 TAVs implanted at node 5 in calcified CV/EV explants. The mean gradient (MG), effective orifice area (EOA), peak velocity, regurgitant fraction (RF), geometric orifice area (GOA), leaflet overhang, leaflet pinwheeling, neoskirt height, and frame deformation were evaluated.

RESULTS

CV/EV explants were calcified and stenotic. Following S3 implantation, the MG and peak velocity decreased. As per the ISO standard, all S3 implants showed adequate EOA, and 3 out of 4 had an RF within the accepted value (<20%). CV/EV leaflet overhang ranged from 25-37%. Calcified leaflets remained stationary throughout the cardiac cycle (difference <9%) and were not pinned in a manner that constrained S3 systolic flow or appeared to prevent selective frame cannulation. The downstream CV/EV GOA was larger than the upstream S3 GOA during systole. S3 frame underexpansion was seen, resulting in leaflet pinwheeling (range 13-30%). Above the neoskirt, calcium protrusion was observed in contact with the S3 leaflets.

CONCLUSIONS

S3 implantation at node 5 in calcified CV/EV valves resulted in satisfactory hydrodynamic performance in most configurations tested with stable leaflet overhang throughout the cardiac cycle. The long-term implications of S3 underexpansion, leaflet pinwheeling, and calcium protrusion require future studies.

Для цитирования: Meier David, Nigade Anish, Lai Althea, Dorman Kyle, Gill Hacina, Javani Shahnaz, Akodad Mariama, Wood David A, Rogers Toby, Puri Rishi, Allen Keith B, Chhatriwalla Adnan K, Reardon Michael J, Tang Gilbert H L, Bapat Vinayak N, Webb John G, Fukuhara Shinichi, Sellers Stephanie L. Redo-TAVI with the SAPIEN 3 valve in degenerated calcified CoreValve/Evolut explants.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2024. DOI: 10.4244/eij-d-24-00619.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.