MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/neurintsurg-2014-011306

Медицинская библиотека · Journal of neurointerventional surgery · 2014

Micro Vascular Plug (MVP)-assisted vessel occlusion in neurovascular pathologies: technical results and initial clinical experience.

Beaty Narlin B, Jindal Gaurav, Gandhi Dheeraj

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Deconstructive approaches may be necessary to treat a variety of neurovascular pathologies. Recently, a new device has become available for endovascular arterial occlusion that may have unique applications in neurovascular disease. The Micro Vascular Plug (MVP, Reverse Medical, Irvine, California, USA) has been designed for vessel occlusion through targeted embolization.

PURPOSE: To report the results from our initial experience with eight consecutive patients in whom the MVP was used to achieve endovascular occlusion of an artery in the head and neck.

METHODS

Eight consecutive patients treated over a nine-month period were included. The patients' radiographic and electronic medical records were retrospectively reviewed. Specifically demographic information, clinical indication, site of arterial occlusion, size of MVP, time to vessel occlusion, clinical complications, use of other secondary embolic agents, and clinical outcome were recorded. Follow-up information when available is presented.

RESULTS

The MVP was used in eight patients for the treatment of neurovascular disease. Indications for treatment included post-traumatic head/neck bleeding (n=3), carotid-cavernous fistula (1), vertebral-vertebral fistula (1), giant fusiform vertebral aneurysm (1), stump-emboli after carotid dissection (1), and iatrogenic vertebral artery penetrating injury (1). One device was used in five patients, two in two patients, and one patient with extensive vertebral-vertebral venous fistula required three plugs to effectively trap the fistula from proximal and distal aspects. Vessel occlusion was obtained in <2 min in each case and there were no procedural complications. Four patients were followed up and no incidence of plug migration or vessel recanalization was seen.

CONCLUSIONS

To the best of our knowledge, this is the first series reporting the use of MVP in neurovascular disease. Use of this device may be associated with shorter procedural times and cost savings in comparison with the use of microcoils for vessel occlusion. Our experience shows that MVP can have unique applications in neurovascular pathologies and it complements other occlusive devices.

Для цитирования: Beaty Narlin B, Jindal Gaurav, Gandhi Dheeraj. Micro Vascular Plug (MVP)-assisted vessel occlusion in neurovascular pathologies: technical results and initial clinical experience.. Journal of neurointerventional surgery. 2014. DOI: 10.1136/neurintsurg-2014-011306.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.