Evaluation of the Kaira COVID-19/Flu/RSV Detection Kit for detection of SARS-CoV-2, influenza A/B, and respiratory syncytial virus: A comparative study with the PowerChek SARS-CoV-2, influenza A&B, RSV Multiplex Real-time PCR Kit.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kim Tae Yeul, Bae Go Eun, Kim Ji-Youn, Kang Minhee, Jang Ja-Hyun, Huh Hee Jae, Chung Doo Ryeon, Lee Nam Yong
Аннотация
BACKGROUND: Co-circulation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and other respiratory viruses, such as influenza and respiratory syncytial virus (RSV), can be a severe threat to public health. The accurate detection and differentiation of these viruses are essential for clinical laboratories. Herein, we comparatively evaluated the performance of the Kaira COVID-19/Flu/RSV Detection Kit (Kaira; Optolane, Seongnam, Korea) for detection of SARS-CoV-2, influenza A and B, and RSV in nasopharyngeal swab (NPS) specimens with that of the PowerChek SARS-CoV-2, Influenza A&B, RSV Multiplex Real-time PCR Kit (PowerChek; Kogene Biotech, Seoul, Korea).
METHODS: A total of 250 archived NPS specimens collected for routine clinical testing were tested in parallel by the Kaira and PowerChek assays. RNA standards were serially diluted and tested by the Kaira assay to calculate the limit of detection (LOD).
RESULTS: The positive and negative percent agreements between the Kaira and PowerChek assays were as follows: 100% (49/49) and 100% (201/201) for SARS-CoV-2; 100% (50/50) and 99.0% (198/200) for influenza A; 100% (50/50) and 100% (200/200) for influenza B; and 100% (51/51) and 100% (199/199) for RSV, respectively. The LODs of the Kaira assay for SARS-CoV-2, influenza A and B, and RSV were 106.1, 717.1, 287.3, and 442.9 copies/mL, respectively.
CONCLUSIONS: The Kaira assay showed comparable performance to the PowerChek assay for detection of SARS-CoV-2, influenza A and B, and RSV in NPS specimens, indicating that the Kaira assay could be a useful diagnostic tool when these viruses are co-circulating.
METHODS: A total of 250 archived NPS specimens collected for routine clinical testing were tested in parallel by the Kaira and PowerChek assays. RNA standards were serially diluted and tested by the Kaira assay to calculate the limit of detection (LOD).
RESULTS: The positive and negative percent agreements between the Kaira and PowerChek assays were as follows: 100% (49/49) and 100% (201/201) for SARS-CoV-2; 100% (50/50) and 99.0% (198/200) for influenza A; 100% (50/50) and 100% (200/200) for influenza B; and 100% (51/51) and 100% (199/199) for RSV, respectively. The LODs of the Kaira assay for SARS-CoV-2, influenza A and B, and RSV were 106.1, 717.1, 287.3, and 442.9 copies/mL, respectively.
CONCLUSIONS: The Kaira assay showed comparable performance to the PowerChek assay for detection of SARS-CoV-2, influenza A and B, and RSV in NPS specimens, indicating that the Kaira assay could be a useful diagnostic tool when these viruses are co-circulating.