DOI 10.1111/nmo.14783

Медицинская библиотека · Neurogastroenterology and motility · 2024

Clinical validation of the VIPUN™ gastric monitoring system versus manometry for the evaluation of gastric motility.

Raymenants Karlien, Huang I-Hsuan, Goelen Nick, Janssen Pieter, Van Tichelen Nico, Burton Duane, Tack Jan

СтатьяВалидационное исследованиеСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Gastrointestinal dysmotility is frequently suspected in patients with gastroparesis, functional dyspepsia, and ileus, and in the intensive care unit. Monitoring of gastric motility in clinical practice remains challenging. A novel technology was developed to meet the medical need for a widely available bedside tool to monitor gastric motility continuously. The VIPUN™ Gastric Monitoring System (GMS) comprises a nasogastric feeding tube with intragastric balloon to allow for measuring gastric contractions.

AIMS: To compare the performance of the VIPUN GMS versus a reference technique (manometry).

METHODS

In this validation study in healthy subjects, the investigational catheter and a solid-state manometry catheter were placed in the stomach concomitantly. Motility was recorded for 2.5 h: 2 h in a fasting state, followed by a 400-kcal liquid meal, and monitoring of the fed state for the remaining half hour. The performance of both systems was compared by automated recognition and manual identification of the contractile activity. Data are presented as mean (standard deviation).

KEY

RESULTS

The analysis set comprised 13 healthy subjects (6 women, age: 27.5 (8.1) years, BMI: 22.2 (2.46) kg/m2). Automatically-recognized contractility was strongly correlated between the two techniques (endpoint: contraction duration; Spearman ρ = 0.96, p < 0.001). A correlation was also observed between the number of individual contractions identified by expert gastroenterologists on both technologies independently (ρ = 0.71, p = .007) and between the contractions identified by the experts and by the GMS software (ρ = 0.87, p = 0.001). No serious or unanticipated adverse events occurred.

CONCLUSIONS & INFERENCES: The observed strong correlations with the gold standard, manometry, validate the performance of the VIPUN GMS as a gastric monitoring system.

Для цитирования: Raymenants Karlien, Huang I-Hsuan, Goelen Nick, Janssen Pieter, Van Tichelen Nico, Burton Duane, Tack Jan. Clinical validation of the VIPUN™ gastric monitoring system versus manometry for the evaluation of gastric motility.. Neurogastroenterology and motility. 2024. DOI: 10.1111/nmo.14783.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.